Single-incision and single-element array electrode to lower the defibrillation threshold.
Abstract
Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and the use of biphasic shocks. A single-element subcutaneous array electrode was implanted in 2 patients with a high defibrillation threshold. The array electrode was implanted through the same infraclavicular incision that was used for implantation of the transvenous lead. The defibrillation threshold decreased from 30 J to 15 J and from 24 J to 9 J with the subcutaneous array electrode.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
V Kühlkamp, K Khalighi, V Dörnberger, G Ziemer. 1997. Single-incision and single-element array electrode to lower the defibrillation threshold.. https://doi.org/10.1016/s0003-4975(97)00811-4