PubMed · 9364592
Simultaneous PET and MR imaging.
Abstract
We have developed a prototype PET detector which is compatible with a clinical MRI system to provide simultaneous PET and MR imaging. This single-slice PET system consists of 48 2 x 2 x 10 mm3 LSO crystals in a 38 mm diameter ring configuration that can be placed inside the receiver coil of the MRI system, coupled to three multi-channel photomultipliers housed outside the main magnetic field via 4 m long and 2 mm diameter optical fibres. The PET system exhibits 2 mm spatial resolution, 41% energy resolution at 511 keV and 20 ns timing resolution. Simultaneous PET and MR phantom images were successfully acquired.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y Shao, S R Cherry, K Farahani, K Meadors, S Siegel, R W Silverman, P K Marsden. 1997. Simultaneous PET and MR imaging.. https://doi.org/10.1088/0031-9155%2F42%2F10%2F010