Vestibular stimulation by multichannel cochlear implants.

Abstract

The recipient of a Nucleus 22 multichannel cochlear implant began to experience severe vestibular stimulation related to the implant. This patient's experience initiated a study with the objective of determining the frequency of implant-related vestibulo-ocular stimulation. Subjects consisted of 17 randomly selected patients who use cochlear implants. Included in the study were 14 Nucleus 22 and three Med-El Combi 40-devices. Stimulation of the implants was performed both by individual channel and with sound field broad-band 80-dB noise using the users' normal device settings. Eye movements were monitored with infrared videonystagmography. Only one subject, who used a Med-El Combi 40, showed a consistent and strong ocular response to cochlear stimulation but had no subjective symptoms. The authors conclude that vestibulo-ocular activation is possible with multichannel cochlear implants but is infrequent and may not be clinically significant.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M L Bance, M O'Driscoll, E Giles, R T Ramsden. 1998. Vestibular stimulation by multichannel cochlear implants.. https://doi.org/10.1097/00005537-199802000-00025

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Pseudospontaneous activity: stochastic independence of auditory nerve fibers with electrical stimulation.

We describe a novel signal processing strategy for cochlear implants designed to emphasize stochastic independence across the excited neural population. The strategy is based on the observation that high rate pulse trains may produce random spike patterns in auditory nerve fibers that are statistically similar to those produced by spontaneous activity in the normal cochlea. We call this activity 'pseudospontaneous'. A supercomputer-based computational model of a population of auditory nerve fibers suggests that different average rates of pseudospontaneous activity can be created by varying the stimulus current of a fixed-amplitude, high-rate pulse train, e.g. 5000 pps. Electrically-evoked compound action potentials recorded in a human cochlear implant subject are consistent with the hypothesis that such a stimulus can desynchronize the fiber population. This desynchronization may enhance neural representation of temporal detail and dynamic range with a cochlear implant and eliminate a major difference between acoustic and electric hearing.

Cochlear Implants

The influence of ionizing radiation on the CLARION 1.2 cochlear implant during radiation therapy.

OBJECTIVE: This study aimed to determine the maximum dose of radiation the CLARION 1.2 cochlear implant can withstand safely. INTRODUCTION: Cochlear implants restore functional hearing to patients with sensorineural deafness. Because some patients may need radiation therapy, it is important to investigate the influence of ionizing radiation on cochlear implant function. METHODS: This study tested the function of four CLARION 1.2 implants (Advanced Bionics, Sylmar, CA, U.S.A.) after varying radiation treatments with gamma rays. The first implant received a cumulative dosage of 69 Gy over nine treatments (single doses between 0.1-30 Gy). The second was irradiated with a total of 90 Gy, receiving three treatments of 30 Gy each. The third and fourth received doses more typical of patient therapy (i.e., 2 Gy) approximately 30 times, for a cumulative dosage of approximately 60 Gy. Implant function was tested after every treatment; the CLARION implant incorporates a back-telemetry system, allowing impedance and current output testing. RESULTS: Despite the type of treatment, the results were quite consistent: difficulties in function occurred when the cumulative dosage inside the implant was approximately 60 Gy. The first implant recovered completely and the second recovered partially. DISCUSSION: The CLARION 1.2 cochlear implant seems to safely withstand approximately 60 Gy of radiation before experiencing functional difficulties. In a clinical situation, the implant would not likely be in the target volume irradiated, and thus the patient's therapeutic cumulative dosage might be higher.

Cochlear Implants