PubMed1999
An image processing strategy is presented that assures very similar soft-copy presentation on diagnostic workstations of a picture archiving and communication system (PACS) over the lifetime of an image file and simultaneously provides efficient work-flow. The strategy is based on rigid partitioning of image processing into application- and display-device-specific processing. Application-specific processing is optimized for a reference display system. A description of this system is attached to the file header of the application-specifically processed image which is stored in the PACS. Every diagnostic display system automatically reproduces the image quality for which the application-specific processing was optimized by adjusting its properties by display-system-specific processing so that the system becomes effectively equal to the reference display system.