[Window width as a dosage-relevant factor in high-contrast structures in CT].

Abstract

PURPOSE: To establish the relationship between window width and dose to be applied in low-dose high-resolution (HR) computed tomography (CT). MATERIAL AND METHODS: Low-dose HR CT-scans of the petrous bone displayed with different window values were analyzed for identification of osseous details. For two homogeneous phantoms, standard deviation of CT-numbers were measured in order to calculate the fraction of pixels not displayed within the correct grey level. RESULTS: The broadest window used (4000 HU) allowed the best distinction of osseous structures. Standard deviation of CT-numbers in the phantoms varied between 28.5 and 43.2 (mean = 33.6) HU. Thus, only 26.6% of all pixels are displayed in the correct grey level for a 1000 HU-window, but 82.3% for a 4000 HU-window. CONCLUSION: When using low doses and HR-reconstruction algorithms, large window widths allow for an optimal assessment of high-contrast structures. Under these conditions, even high standard deviations of the CT-numbers will not compromise image evaluation. Due to the restricted number of grey-levels that can be distinguished by the human eye, image noise is compressed into a smaller number of discernable grey-levels. Because of the inverse square root function between dose to be applied and window width used (derived in the article), an even minor enlargement of window permits considerable reductions of patient exposure while the discernable image noise remains constant.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Husstedt, M Prokop, H Becker. 1998. [Window width as a dosage-relevant factor in high-contrast structures in CT].. https://doi.org/10.1055/s-2007-1015198

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2 µm × 2 µm resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗