Reliability of closed double helix electrode for functional electrical stimulation.

Abstract

The reliability of a closed double helix electrode in the lower limbs was studied. This electrode is an implanted intramuscular electrode and is used for a totally implantable functional electrical stimulation system. Eighty electrodes were evaluated retrospectively with a mean period of 15 months. The total implant time was 1222 electrode months. The cumulative proportion surviving was 0.934 at 6 months, 0.855 at 1 year, 0.765 at 2 years, and 0.730 after 30 months. Fifteen of 80 electrodes failed, seven showed increasing electrode impedance, and eight had undesirable changes in recruitment. Of the failed electrodes, 2/3 failed during the first 10 months. The reliability was 0.91 at 6 months and 0.80 at 1 year after implantation in all muscle groups. The closed double helix electrode displayed an increased reliability when compared with the open double helix electrode at 6 months, and an equivalent reliability as compared with the electrodes developed by Handa and colleagues at 6 months and 1 year, using the chi squared test for independence. This study suggests that the closed double helix electrode has an acceptable reliability and can be used as a part of a totally implantable functional electrical stimulation system.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Kagaya, M Sharma, G Polando, E B Marsolais. 1998. Reliability of closed double helix electrode for functional electrical stimulation.. https://pubmed.ncbi.nlm.nih.gov/9577430/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Modeling biological motor control for human locomotion with functional electrical stimulation.

This paper develops a novel control system for functional electrical stimulation (FES) locomotion, which aims to generate normal locomotion for paraplegics via FES. It explores the possibility of applying ideas from biology to engineering. The neural control mechanism of the biological motor system, the central pattern generator, has been adopted in the control system design. Some artificial control techniques such as neural network control, fuzzy logic, control and impedance control are incorporated to refine the control performance. Several types of sensory feedback are integrated to endow this control system with an adaptive ability. A musculoskeletal model with 7 segments and 18 muscles is constructed for the simulation study. Satisfactory simulation results are achieved under this FES control system, which indicates a promising technique for the potential application of FES locomotion in future.

Electric Stimulation Therapy↗

Complications and troubleshooting of two-stage sacral neuromodulation therapy: a single-institution experience.

OBJECTIVES: With the increasing use of sacral neuromodulation therapy, urologists are faced with postimplantation challenges. The purpose of this study was to identify these events and their causes and management in our large single-institution experience. METHODS: From July 2002 to September 2004, all patients who underwent sacral neuromodulation therapy for refractory bladder conditions were identified. Their medical records were retrospectively evaluated for history, operative intervention, and programming visits. Events after implantation of the implantable pulse generator (IPG) unit were noted. The evaluation, troubleshooting, management, and resolution of events at the last follow-up visit were extracted. RESULTS: A total of 214 patients underwent sacral neuromodulation therapy at our institution. The mean patient age was 53.5 +/- 15.4 years. Of the 214 patients, 161 underwent IPG implantation during a mean follow-up period of 16 months (range 5 to 30). The second-stage explantation and revision rate was 10.5% and 16.1%, respectively. The indications for explantation were infection (8 of 17) and failure to maintain a response (9 of 17). Revisions were done for decreases in response with abnormal (12 of 26) or normal (5 of 26) impedance measurements, IPG site discomfort (4 of 26), draining sinus at the IPG site (4 of 26), and lead migration (1 of 26). Equalization of impedance measurements was the most commonly observed impedance abnormality. These were managed by drying fluid from the connection in 4 patients and lead change in the rest. CONCLUSIONS: After IPG implantation, a decline in response may occur. Although some were explanted, most were revised, with most of the revisions functional on follow-up. Familiarity with impedance evaluation and development of algorithms for postimplant management are essential for troubleshooting and maintenance of the device.

Electric Stimulation Therapy↗