Abstract
Several in-the-ear (ITE) and behind-the-ear (BTE) hearing aids were tested for audible interference at various distances from five types of digital wireless telephones. The interference which takes the form of a buzzing and a static sound was quantified using a calibrated system including a frequency analyzer and a pressure field microphone. The output of the each hearing aid was coupled to the microphone via Tygon tubing and a standard 2 cc coupler. The highest interference-induced sound pressure level (SPL), 122.5 dB, was measured from a BTE hearing aid placed within 2 cm of a transmitting Global System for Mobile Communications (GSM) phone. In this case, interference was detected up to a separation distance of almost 3 m. While all phones tested produced a similar interference level within 2 cm of this hearing aid, interference SPL from the code division multiple access (CDMA)-based system decreased more rapidly with distance than the time division multiple access (TDMA)-based phones tested.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Skopec. 1998. Hearing aid electromagnetic interference from digital wireless telephones.. https://doi.org/10.1109/86.681190
Cite the original work for its findings. Save a collection to share your selection of sources.