PubMed · 9702716
Single-shot, variable flip-angle slice-selective excitation with four gradient-modulated adiabatic half-passage segments.
Abstract
Adiabatic pulses, although useful in generating uniform spin nutation in the presence of inhomogeneous B1 fields, are limited for NMR imaging applications due to the lack of slice-selective excitation capability. Selective excitation techniques using gradient modulation have been introduced; however, present methods require either a minimum of two excitations or eight adiabatic segments. Here, a scheme is presented that allows single-shot, arbitrary flip-angle, and slice-selective excitation with only four adiabatic half-passage segments. The technique is demonstrated via computer simulation and experimental tests on a phantom. Furthermore, issues associated with the implementation of these gradient-modulated adiabatic pulses are discussed.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
E W Hsu, S B Reeder, J R MacFall. 1998. Single-shot, variable flip-angle slice-selective excitation with four gradient-modulated adiabatic half-passage segments.. https://doi.org/10.1002/mrm.1910400220