PubMed · 9739372
[Magnetization transfer contrast (MTC)--which is the most MTC-sensitive MRI sequence?].
Abstract
PURPOSE: This study aimed at evaluating MTC-sensitivity of frequently used MR-sequences. MATERIAL AND METHODS: 7 sequences (T1-, T2-, PD-weighted SE, four gradient-echo sequences) were combined with a 4 element on-resonance MT-Pulse at 0.5 Tesla (T5, Philips medical systems). The measurements were performed at a knee joint and MT-ratios of cartilage, muscle, fat and a copper sulfate phantom were calculated. RESULTS: Proton density weighted gradient and to a lesser degree spin-echo sequences were superior to T1- and T2-weighted sequences in visualising the MT effects. CONCLUSION: It is important to choose an appropriate MR-sequence for MTC experiments. Proton-density gradient echo sequences are recommended for that purpose. T1- and T2-weighted SE sequences should be avoided.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Vahlensieck, J Gieseke, F Träber, H Schild. 1998. [Magnetization transfer contrast (MTC)--which is the most MTC-sensitive MRI sequence?].. https://doi.org/10.1055/s-2007-1015073
Cite the original work for its findings. Save a collection to share your selection of sources.