Comparison of three computerized videokeratoscopy systems with keratometry.

Abstract

PURPOSE: To compare the reproducibility of computerized videokeratoscopy systems by using normal eyes and calibrated objects. METHODS: We evaluated the reproducibility of three commercially available videokeratoscopes [EyeSys, TechnoMed C-Scan, and PAR Corneal Topography System (CTS)] with the manual keratometer (Bausch & Lomb) by using calibration spheres and 10 normal subjects (20 eyes). All videokeratoscopy and keratometer results were obtained by one investigator (R.M.). Each eye and calibration sphere were submitted to 10 serial examinations by using each system. The average K of all points within the central 3.0 mm of the topography systems (central 3.0 mm) was compared with the average K of the manual keratometer. RESULTS: All videokeratoscopy systems correlated well with each other and manual keratometry when accessing aspheric and spherocylinder calibration balls. EyeSys central keratometry clinical results had the strongest correlation with the average keratometry results at 35%, followed by PAR-CTS at 25% and C-Scan at 5%. Among the videokeratoscopy units, EyeSys and PAR-CTS had the strongest correlation at 65%. The correlation between the TechnoMed C-Scan and both the EyeSys and PAR-CTS systems was 25%. There was a statistically significant difference (p < 0.05) between the systems when analyzing the results obtained from clinical subjects. The average keratometry (K) difference of human eyes between videokeratoscopy systems is <0.35 diopters (D) (p < 0.05), which may be clinically significant. The average manual K reading (42.97 D) is statistically significantly flatter (p < 0.05) than each of the videokeratoscopy units (EyeSys = 43.49 D; PAR = 43.48 D; C-Scan = 43.83 D). Comparing the 10 measurements of each eye or calibration object in the same videokeratoscopy system verified that the devices give reproducible results. The average standard deviation (ASD) of the keratometer was 0.10 D. The ASD of the videokeratoscopy units was 0.05 D for the EyeSys, 0.29 D for the PAR-CTS, and 0.31 D for the C-Scan systems. CONCLUSION: Based on this study, we should not assume that the results of different topography systems can be interchanged in clinical studies.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R C Moura, B L Bowyer, S X Stevens, J J Rowsey. 1998. Comparison of three computerized videokeratoscopy systems with keratometry.. https://doi.org/10.1097/00003226-199809000-00010

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Differential tissue-on-tissue lubrication by ophthalmic formulations.

Tissue-on-tissue friction testing was used to determine how instillation of hydrophilic polymer-containing formulations between the "blinking" tissues would compare with lubrication by saline, alone, or an oil-emulsion preparation. Best results were obtained for a formulation that contained active demulcents polyethylene glycol (PEG 400) and propylene glycol (PG), as well as a gellable polymer hydroxypropyl guar (HP-Guar) in a borate-buffered solution, in comparison with hydroxypropylcellulose-containing and carboxymethylcellulose-containing formulations. Superior performance of all the formulations was found for lubricating tissue-on-tissue couples, compared with metal-oxide-to-metal oxide interfaces, or metal oxide-to-tissue interfaces. A reciprocating pin-on-disc type friction/wear test device articulated the intimal faces of preserved human umbilical cord vein segments under increasing loads during simulated continuous "eye-blinking" with addition of increasing weights up to 60 g/cm2, simulating maximal eyelid force on the orbital globe. The tissue-on-tissue couples moved from liquid phase lubrication to boundary lubrication. After residual formulations were rinsed away with saline, persistence of low friction at the highest loads was indicative of formulation substantivity. Human umbilical cord vein segments were utilized in saline-wetted tissue-on-tissue couples that showed variable starting coefficients of friction in the range 0.2-0.4, producing moderate tearing and disruption of the interfacial layers above the medial collagen zone. The best-performing formulations instilled to the tissues pre-wetted with saline apparently reacted separately with each tissue face to produce a lower final and persistent coefficient of friction of about 0.05. Scanning electron microscopy and light microscopy of these guar-modified tissue specimens showed only a few superficial tissue disruptions, and some interphase swelling consistent with polymer uptake. The frictional values for lubricated couples having non-tissue members were considerably higher than the coefficients of friction measured for the similarly lubricated tissue-on-tissue couples, emphasizing the requirement that appropriate simulations are critical to obtaining clinically predictive data.

Cornea↗

Validated prediction model for the development of primary open-angle glaucoma in individuals with ocular hypertension.

OBJECTIVE: To test the validity and generalizability of the Ocular Hypertension Treatment Study (OHTS) prediction model for the development of primary open-angle glaucoma (POAG) in a large independent sample of untreated ocular hypertensive individuals and to develop a quantitative calculator to estimate the 5-year risk that an individual with ocular hypertension will develop POAG. DESIGN: A prediction model was developed from the observation group of the OHTS and then tested on the placebo group of the European Glaucoma Prevention Study (EGPS) using a z statistic to compare hazard ratios, a c statistic for discrimination, and a calibration chi2 for systematic overestimation/underestimation of predicted risk. The 2 study samples were pooled to increase precision and generalizability of a 5-year predictive model for developing POAG. PARTICIPANTS: The OHTS observation group (n = 819; 6.6 years' median follow-up) and EGPS placebo group (n = 500; 4.8 years' median follow-up). TESTING: Data were collected on demographic characteristics, medical history, ocular examination visual fields (VFs), and optic disc photographs. MAIN OUTCOME MEASURE: Development of reproducible VF abnormality or optic disc progression as determined by masked readers and attributed to POAG by a masked end point committee. RESULTS: The same predictors for the development of POAG were identified independently in both the OHTS observation group and the EGPS placebo group-baseline age, intraocular pressure, central corneal thickness, vertical cup-to-disc ratio, and Humphrey VF pattern standard deviation. The pooled multivariate model for the development of POAG had good discrimination (c statistic, 0.74) and accurate estimation of POAG risk (calibration chi2, 7.05). CONCLUSIONS: The OHTS prediction model was validated in the EGPS placebo group. A calculator to estimate the 5-year risk of developing POAG, based on the pooled OHTS-EGPS predictive model, has high precision and will be useful for clinicians and patients in deciding the frequency of tests and examinations during follow-up and advisability of initiating preventive treatment.

Cornea↗

Corneal oxygen uptake: A review of polarographic techniques, applications, and variables.

Factors that influence the polarographic measurement of the oxygen uptake of the cornea are reviewed. These factors include the technique, electrode assembly, oxygen reservoir, membrane material and thickness, oxygen tension of the corneal environment, duration of exposure to environmental conditions and time to application of the probe all influencing measured oxygen uptake rates. Subject factors include lid position, palpebral aperture size, blinking, corneal thickness, and corneal integrity. Contact lens wear influences corneal oxygen uptake, with lens material and design parameters influencing rates obtained both under static (without blinking) and dynamic (with blinking) conditions. Measurement of corneal oxygen uptake rates remains an excellent method to quantify the oxygen supply in contact lens systems that include the contact lens, the tears, and the cornea, in which oxygen flux is influenced by the thickness and diffusion characteristics of each component.

Cornea↗