PubMed · 9877460
Transverse gradient coil with circle current paths.
Abstract
The paper describes a new type of transverse magnetic field gradient coil for applications in MRI and MRS. It uses full circle current paths rather than current arcs and has a particularly simple unit construction of symmetric form. It features a large volume of uniform transverse field gradient, high efficiency and low inductance for rapid switching. A prototype coil has been constructed and evaluated to check computer simulations. Two smaller sets have subsequently been made for NMR microscopy. This coil design may be used for all sizes.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
E R Andrew, M Kempka. 1998. Transverse gradient coil with circle current paths.. https://doi.org/10.1007/bf02592257