PubMed Health⌕ Search

Biomedical subjects

B M Wroblewski

Publications and source records attributed to B M Wroblewski.

At least 145 records · Page 8Linked to original sources

Direction and rate of socket wear in Charnley low-friction arthroplasty.

Twenty-two high-density polyethylene sockets from Charnley low-friction arthroplasties have been studied. Acrylic casts and shadowgraph techniques were used to measure both the real and radiographic rates of wear; these rates showed a significant correlation. In the sagittal plane, nine of the sockets had worn lateral to a line drawn vertically from the centre of curvature of the socket, 12 had worn medial to that line and only one was worn exactly in the line. In the coronal plane, nine sockets had worn in front of a similar vertical line, two behind that line and 11 had worn exactly in the line. Evidence of impingement of the neck of the stem onto the rim of the socket was found in 14 patients; this is considered to be one of the causes of socket loosening. The obvious solution is to reduce the diameter of the neck of the stem from 12.5 mm to 10 mm; provided that it is made of cold-formed, high nitrogen-content stainless steel, this narrower neck is strong enough not to fracture.

Adult↗

Current trends in revision of total hip arthroplasty.

Total hip arthroplasty has brought pain relief and mobility to many patients, but load bearing implants may fail and the resulting problems must be tackled early and with vigour. The primary operation offers the best chance of success, and revision may only ever be considered a salvage procedure. Patients with total hip arthroplasty must be offered facilities for an indefinite follow-up. Revision surgery is not to be undertaken lightly or on an occasional basis, and the lessons learned from such procedures must be used for the benefit of the patients undergoing primary operation.

Hip Dislocation↗

Intramedullary cancellous bone block to improve femoral stem fixation in Charnley low-friction arthroplasty.

In order to improve the fixation of the stem in primary Charnley low-friction arthroplasty, the medullary canal was routinely closed off with a cancellous bone block. A prospective study of 611 consecutive arthroplasties were carried out between 1 and 5 1/2 years (average 2 years and 9 months) after operation. There were two cases of radiological loosening of the stem, both of which could have been avoided. Five other patients showed demarcation of the bone-cement junction at the calcar; two of these are considered to be "at risk" of loosening. There was no evidence of radiological loosening of the stem in 604 cases and the method is recommended for routine use in primary total hip arthroplasty.

Adult↗

Charnley low-friction arthroplasty for the treatment of hips with late complications of femoral neck fractures.

Two hundred five Charnley low-friction arthroplasties (LFA) were performed in 205 patients for late complications of femoral neck fractures. LFA was an excellent procedure for relieving pain and restoring function with a good range of movement. The deep infection rate was 3.9%. A detailed analysis of the patients with infection showed a high incidence of previous superficial wound infection or positive bacteriologic culture swabs obtained from the wound at the time of LFA.

Aged↗

Charnley low-friction arthroplasty in diabetic patients.

A retrospective study of 44 diabetic patients who, between them, had 62 Charnley low-friction arthroplasties, showed a superficial-infection rate of 9.7 per cent and a deep-infection rate of 5.6 per cent. All the operations were carried out in the Charnley clear-air enclosure and prophylactic antibiotics were not used. A statistically significant increase in the overall rate of infection was found in diabetic patients when compared with non-diabetic osteoarthritic patients (P less than 0.001) and rheumatoid patients (P less than 0.01). Hence it is suggested that in diabetic patients there may well be a place for prophylactic antibiotics in hip replacement surgery in addition to the use of the clean-air enclosure.

Adult↗

Bacteriological swabs in hip surgery.

A prospective trial was carried out in 70 patients undergoing elective total hip replacement, in order to correlate bacterial contaminants in the wound at operation with the subsequent clinical results. A new method of collecting bacteria via a swab left in the wound for the duration of the operation proved to be as effective as the conventional swabbing technique in patients with known previous hip sepsis who, following revision surgery, showed evidence of further infection.

Bacteriological Techniques↗

Calcar femorale.

Explore the source record for details and available documents.

Femur Neck↗

Charnley low-friction arthroplasty in patients with psoriasis.

A retrospective study of 38 patients with established psoriasis treated by 55 Charnley low-friction arthroplasties showed a superficial infection rate of 9.1% and a deep infection rate of 5.5%. All of the operations were carried out in the Charnley clean-air enclosure, without prophylactic antibiotic therapy. The incidence of postoperative sepsis in psoriatic patients was significantly higher than in patients with osteoarthritis (p less than 0.001) and rheumatoid arthritis (p less than 0.05). Hence, prophylactic antibiotic therapy appears beneficial in psoriatic patients.

Adult↗

Fractured stem in total hip replacement. A clinical review of 120 cases.

Clinical review of 120 cases of fractured stem of the now obsolete Charnley "flat back" has shown an incidence of 1.15 per cent. The period of risk was the 11 years from the time of surgery and 97.5 per cent fractured during that time. Before the fractured occurred 87.8 per cent had normal hip function. There was a linear relationship between the patients' weight and the time of the fracture. Patients of comparable weight, and referred from other units, fractured the stem earlier than patients from our unit. Radiographic evidence of failure of the femoral stem fixation, within 1 year of surgery, was found in 77.2 per cent of the hips where serial radiographs were available. This group fractured the stem, on average, 17.3 months earlier. Loss of calcar height was present in 73.3 per cent and was a technical problem at revision. Study of varus/valgus orientation of the stem did not bear out the commonly held opinion; in patients of comparable weight valgus stems fractured earlier than varus stems. Mechanical loosening of the socket in the group was 12.9 per cent. Detachment of the trochanter following revision was 12.1 per cent and the dislocation rate 5.2 per cent. The findings suggest that the surgical technique, as regards preparation of the medullary canal and the cement injection, and the patients' weight, are the most important factors in the complication under study.

Aged↗

Recurrence of ectopic ossification after excision in Charnley Low Friction Arthroplasty.

In 23 hips with a Charnley Low Friction Arthroplasty (L.F.A.), revised for mechanical reasons, a planimeter was used to study areas of ectopic ossification and their recurrence after excision. In two cases some bone resorption had taken place. In the remaining 21 hips the average area of recurrence was greater than the average area excised. There was no correlation between the time elapsed and the extent of ossification; once established, the ossification did not seem to progress. Revision of L.F.A. for excision of ectopic bone is not indicated.

Adult↗

Revision surgery in total hip arthroplasty: surgical technique and results.

Revisions in failed total hip arthroplasty have produced some good results. In cases of stem fracture, the problem is lack or loss of proximal support in the presence of distal fixation of the stem. This aspect must be borne in mind during primary and revision surgery. Routine use of long intramedullary stems has little to commend it. Proximal support of the stem is essential. In revision for deep sepsis, antibiotic-loaded cement has increased the success rate. Sound component fixation is, however, of paramount importance. If success is to be achieved in revision surgery, the patients at risk must be carefully followed up by serial radiographs. The timing of the revision will depend on awareness of the problem rather than the patients' symptoms. This, at times, may mean revising early and for radiographic changes alone if the quality of the bone stock is to be preserved, and if the chances of success are to be improved.

Bacterial Infections↗

Bilateral Charnley arthroplasty as a single procedure. A report on 400 patients.

A series of 400 patients (average age 57.7 years) with bilateral low-friction arthroplasties of the hip has been reviewed. Comparison was made with similar unilateral operations. The results showed that the bilateral arthroplasties can be carried out as a single operation without higher risk of local or general complications. However, this excludes the incidence of pulmonary embolism which is higher for the bilateral replacements.

Adult↗

Revision of the Charnley low-friction arthroplasty for recurrent or irreducible dislocation.

Twenty-one cases of the Charnley low-friction arthroplasty were revised because of recurrent or irreducible dislocation. Three main reasons for this revision, usually in combination, were identified: loss of the abductor mechanism due to trochanteric detachment; shortening of the limb due to high placement of the socket or low section of the femoral neck; and malorientation of the components. Sixteen patients had no further problems after revision. One patient had a single dislocation, four had more than one dislocation although they managed to cope with them and did not require a further revision.

Adult↗

Blood loss and operation time in the Charnley low friction arthroplasty.

Review of 181 patients undergoing 194 Charnley low friction arthroplasties has shown that the average time taken for primary surgery was 65 minutes. The operative blood loss was 192 ml and postoperative drainage 437 ml. In conversion from failed hip surgery the blood loss was 260 ml at operation with an average drainage of 610 ml. The operative time was 87 minutes on average. Revision of failed hip replacement was carried out in an average time of 101 minutes. Average operative blood loss was 510 ml and postoperative drainage 680 ml. The findings show that criticism suggesting that detachment of the greater trochanter during total hip replacement increases operative time and blood loss is unfounded.

Drainage↗

The mechanism of fracture of the femoral prosthesis in total hip replacement.

Examination of the fractured stems of 70 Charnley, flat back, femoral prostheses suggested that the causative mechanism was a torisonal loading of the proximal portion on the distal part of the stem. The main loading on the hip is applied from an anterosuperior direction. The square corners of the stem and its implantation in retroversion predispose to fracture.

Hip Fractures↗

Wear of high-density polyethylene on bone and cartilage.

Wear of high-density polyethylene on bone and cartilage has resulted in a large volume of plastic particles being shed into the two knees and two hips studied. The giant-cell foreign-body reaction of the synovium may not be sufficient to cope with the amount of debris presented and the destruction of the endosteal bone in one hip, caused by the wear particles and movement of the prosthesis, has made revision impossible. Articulation of high-density polyethylene against bone or cartilage either by design or by the failure of alignment of the component must be avoided.

Bone and Bones↗