PubMed Health⌕ Search

Biomedical subjects

Christopher M Collins

Publications and source records attributed to Christopher M Collins.

At least 19 recordsLinked to original sources

MR thermometry characterization of a hyperthermia ultrasound array designed using the k-space computational method.

BACKGROUND: Ultrasound induced hyperthermia is a useful adjuvant to radiation therapy in the treatment of prostate cancer. A uniform thermal dose (43 degrees C for 30 minutes) is required within the targeted cancerous volume for effective therapy. This requires specific ultrasound phased array design and appropriate thermometry method. Inhomogeneous, acoustical, three-dimensional (3D) prostate models and economical computational methods provide necessary tools to predict the appropriate shape of hyperthermia phased arrays for better focusing. This research utilizes the k-space computational method and a 3D human prostate model to design an intracavitary ultrasound probe for hyperthermia treatment of prostate cancer. Evaluation of the probe includes ex vivo and in vivo controlled hyperthermia experiments using the noninvasive magnetic resonance imaging (MRI) thermometry. METHODS: A 3D acoustical prostate model was created using photographic data from the Visible Human Project. The k-space computational method was used on this coarse grid and inhomogeneous tissue model to simulate the steady state pressure wavefield of the designed phased array using the linear acoustic wave equation. To ensure the uniformity and spread of the pressure in the length of the array, and the focusing capability in the width of the array, the equally-sized elements of the 4 x 20 elements phased array were 1 x 14 mm. A probe was constructed according to the design in simulation using lead zerconate titanate (PZT-8) ceramic and a Delrin plastic housing. Noninvasive MRI thermometry and a switching feedback controller were used to accomplish ex vivo and in vivo hyperthermia evaluations of the probe. RESULTS: Both exposimetry and k-space simulation results demonstrated acceptable agreement within 9%. With a desired temperature plateau of 43.0 degrees C, ex vivo and in vivo controlled hyperthermia experiments showed that the MRI temperature at the steady state was 42.9 +/- 0.38 degrees C and 43.1 +/- 0.80 degrees C, respectively, for 20 minutes of heating. CONCLUSION: Unlike conventional computational methods, the k-space method provides a powerful tool to predict pressure wavefield in large scale, 3D, inhomogeneous and coarse grid tissue models. Noninvasive MRI thermometry supports the efficacy of this probe and the feedback controller in an in vivo hyperthermia treatment of canine prostate.

Acoustics↗

Consideration of magnetically-induced and conservative electric fields within a loaded gradient coil.

We present a method to calculate the electric (E)-fields within and surrounding a human body in a gradient coil, including E-fields induced by the changing magnetic fields and "conservative" E-fields originating with the scalar electrical potential in the coil windings. In agreement with previous numerical calculations, it is shown that magnetically-induced E-fields within the human body show no real concentration near the surface of the body, where nerve stimulation most often occurs. Both the magnetically-induced and conservative E-fields are shown to be considerably stronger just outside the human body than inside it, and under some circumstances the conservative E-fields just outside the body can be much larger than the magnetically-induced E-fields there. The order of gradient winding and the presence of conductive RF shield can greatly affect the conservative E-field distribution in these cases. Though the E-fields against the outer surface of the body are not commonly considered, understanding gradient E-fields may be important for reasons other than peripheral nerve stimulation (PNS), such as potential interaction with electrical equipment.

Computer Simulation↗

Exploring the limits of RF shimming for high-field MRI of the human head.

Several methods have been proposed for overcoming the effects of radiofrequency (RF) magnetic field inhomogeneity in high-field MRI. Some of these methods rely at least in part on the ability to independently control magnitude and phase of different drives in either one multielement RF coil or in different RF coils in a transmit array. The adjustment of these drive magnitudes and phases alone to create uniform RF magnetic (B(1)) fields has been called RF shimming, and has certain limits at every frequency as dictated by possible solutions to Maxwell's equations. Here we use numerical calculations to explore the limits of RF shimming in the human head. We found that a 16-element array can effectively shim a single slice at frequencies up to 600 MHz and the whole brain at up to 300 MHz, while an 80-element array can shim the whole brain at up to 600 MHz.

Computer Simulation↗

Central brightening due to constructive interference with, without, and despite dielectric resonance.

PURPOSE: To aid in discussion about the mechanism for central brightening in high field magnetic resonance imaging (MRI), especially regarding the appropriateness of using the term dielectric resonance to describe the central brightening seen in images of the human head. MATERIALS AND METHODS: We present both numerical calculations and experimental images at 3 T of a 35-cm-diameter spherical phantom of varying salinity both with one surface coil and with two surface coils on opposite sides, and further numerical calculations at frequencies corresponding to dielectric resonances for the sphere. RESULTS: With two strategically placed surface coils it is possible to create central brightening even when one coil alone excites an image intensity pattern either bright on one side only or bright on both sides with central darkening. This central brightening can be created with strategic coil placement even when the resonant pattern would favor central darkening. Results in a conductive sample show that central brightening can similarly be achieved in weakly conductive dielectric materials where any true resonances would be heavily damped, such as in human tissues. CONCLUSION: Constructive interference and wavelength effects are likely bigger contributors to central brightening in MR images of weakly conductive biological samples than is true dielectric resonance.

Electromagnetic Fields↗

Hexagonal zero mode TEM coil: a single-channel coil design for imaging multiple small animals.

A novel hexagonal coil design for simultaneous imaging of multiple small animals is presented. The design is based on a coaxial cavity and utilizes the magnetic field formed between two coaxial conductors with hexagonal cross-sections. An analytical solution describing the B(1) field between conductors of the hexagonal coil was found from the Biot-Savart law. Both experimental results and analytical calculations showed a variation in the B(1) field within the imaging region of less than 10%. Numerical calculations predicted approximately 35% improvement in B(1) field homogeneity with the hexagonal coil design compared to a cylindrical coaxial cavity design. The experimentally-measured signal-to-noise ratio (SNR) of the hexagonal coil loaded with six 50-mM phantoms was only 4-5% lower than that of a single parallel plate resonator loaded with one phantom. In vivo spin-echo (SE) images of six 7-day-old rat pups acquired simultaneously demonstrated sufficient SNR for microimaging. The construction scheme of the coil, simple methods for tuning and matching, and an anesthesia device and animal holder designed for the coil are described. The hexagonal coil design utilizes a single receiver and allows for simultaneous imaging of six small animals with no significant compromise in SNR.

Animals↗

Strategies for shimming the breast.

There is evidence in the literature indicating a significant static field inhomogeneity in the human breast. A nonhomogenous field results in line broadening and frequency shifts in MRS and can cause intensity loss and spatial errors in MRI. Thus, there is a clear rationale for determining the regional variations in the static field homogeneity in the breast and providing strategies to correct them. Herein, the nature and extent of the static magnetic field at 3 T were measured in central planes of the human breast using both phase maps and multivoxel MRS techniques. In addition, the effect of first- and high-order shimming and of spatial saturation pulses on the static field inhomogeneity was evaluated. Both the theoretical and the measured field were found to be primarily linear in nature, with a reduction of 300 Hz from the nipple to the chest wall. First-order shimming reduced this inhomogeneity by 65%. Interestingly, the combination of spatial saturation pulses and first-order shimming was more effective than high-order shim alone. Since many clinical scanners do not have either higher-order shim or automated higher shimming algorithms that work in the presence of fat, the suggested combination provides an effective means to correct inhomogeneities in the breast.

Adult↗

B(1) destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil.

RF behavior in the human head becomes complex at ultrahigh magnetic fields. A bright center and a weak periphery are observed in images obtained with volume coils, while surface coils provide strong signal in the periphery. Intensity patterns reported with volume coils are often loosely referred to as "dielectric resonances," while modeling studies ascribe them to superposition of traveling waves greatly dampened in lossy brain tissues, raising questions regarding the usage of this term. Here we address this question experimentally, taking full advantage of a transceiver coil array that was used in volume transmit mode, multiple receiver mode, or single transmit surface coil mode. We demonstrate with an appropriately conductive sphere phantom that destructive interferences are responsible for a weak B(1) in the periphery, without a significant standing wave pattern. The relative spatial phase of receive and transmit B(1) proved remarkably similar for the different coil elements, although with opposite rotational direction. Additional simulation data closely matched our phantom results. In the human brain the phase patterns were more complex but still exhibited similarities between coil elements. Our results suggest that measuring spatial B(1) phase could help, within an MR session, to perform RF shimming in order to obtain more homogeneous B(1) in user-defined areas of the brain.

Algorithms↗

Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies.

Image inhomogeneity related to high radiofrequencies is one of the major challenges for high field imaging. This inhomogeneity can be thought of as having 2 radiofrequency-field related contributors: the transmit field distribution and the reception field distribution. Adjusting magnitude and phase of currents in elements of a transmit array can significantly improve flip angle homogeneity at high field. Effective application of some well-known parallel imaging and other receive array post-processing methods removes receptivity patterns from the intensity distribution in the final image, though noise then becomes a function of position in the final image. Here simulations are used to show that, assuming high signal-to-noise ratio, very homogeneous images in the human head can be acquired with the combination of transmit arrays and some receive array reconstruction methods at frequencies as high as 600 MHz.

Algorithms↗

Model of local temperature changes in brain upon functional activation.

Experimental results for changes in brain temperature during functional activation show large variations. It is, therefore, desirable to develop a careful numerical model for such changes. Here, a three-dimensional model of temperature in the human head using the bioheat equation, which includes effects of metabolism, perfusion, and thermal conduction, is employed to examine potential temperature changes due to functional activation in brain. It is found that, depending on location in brain and corresponding baseline temperature relative to blood temperature, temperature may increase or decrease on activation and concomitant increases in perfusion and rate of metabolism. Changes in perfusion are generally seen to have a greater effect on temperature than are changes in metabolism, and hence active brain is predicted to approach blood temperature from its initial temperature. All calculated changes in temperature for reasonable physiological parameters have magnitudes <0.12 degrees C and are well within the range reported in recent experimental studies involving human subjects.

Body Temperature↗

Effect of gender on in vivo cartilage magnetic resonance imaging T2 mapping.

PURPOSE: To determine if gender is a significant variable for in vivo magnetic resonance imaging (MRI) T2-mapping of knee articular cartilage in young asymptomatic volunteers. MATERIALS AND METHODS: Cartilage MRI T2 mapping was performed in a young healthy population consisting of seven male and 10 female volunteers, 22 to 29 years of age. High-resolution in vivo T2 maps were obtained of patellar, tibial, and weight-bearing femoral articular cartilage. Spatial dependency of cartilage T2 between groups was evaluated through a comparison of cartilage T2 as a function of normalized distance from bone. RESULTS: Bulk cartilage T2 values were similar at all three anatomic sites, and between male and female volunteers. All volunteers demonstrated similar spatial variation in cartilage MRI T2 values, with a minimum located in the radial zone and increasing T2 values toward the articular surface. There was no difference in spatial dependency of cartilage T2 between males and females. CONCLUSION: In young, healthy volunteers, the magnitude and spatial dependency of cartilage T2 does not differ with gender.

Adult↗

Temperature and SAR calculations for a human head within volume and surface coils at 64 and 300 MHz.

PURPOSE: To examine relationships between specific energy absorption rate (SAR) and temperature distributions in the human head during radio frequency energy deposition in MRI. MATERIALS AND METHODS: A multi-tissue numerical model of the head was developed that considered thermal conductivity, heat capacity, perfusion, heat of metabolism, electrical properties, and density. Calculations of SAR and the resulting temperature increase were performed for different coils at different frequencies. RESULTS: Because of tissue-dependent perfusion rates and thermal conduction, there is not a good overall spatial correlation between SAR and temperature increase. When a volume coil is driven to induce a head average SAR level of either 3.0 or 3.2 W/kg, it is unlikely that a significant temperature increase in the brain will occur due to its high rate of perfusion, although limits on SAR in any 1 g of tissue in the head may be exceeded. CONCLUSION: Attempts to ensure RF safety in MRI often rely on assumptions about local temperature from local SAR levels. The relationship between local SAR and local temperature is not, however, straightforward. In cases where high SAR levels are required due to pulse sequence demands, calculations of temperature may be preferable to calculations of SAR because of the more direct relationship between temperature and safety.

Equipment Safety↗

Effects of end-ring/shield configuration on homogeneity and signal-to-noise ratio in a birdcage-type coil loaded with a human head.

We modeled four different end-ring/shield configurations of a birdcage coil to examine their effects on field homogeneity and signal-to-noise ratio (SNR) at 64 MHz and 125 MHz. The configurations are defined as: 1) conventional: a conventional cylindrical shield; 2) surrounding shield: a shield with annular extensions to closely shield the end rings; 3) solid connection: a shield with annular extensions connected to the rungs; and 4) thin wire connection: a shield with thin wires connected to the rungs. At both frequencies, the coil with conventional end-ring/shield configuration produces the most homogeneous RF magnetic (B1) field when the coil is empty, but produces the least homogeneous B1 field when the coil is loaded with a human head. The surrounding shield configuration results in the most homogeneous B1 and highest SNR in the coil loaded with the human head at both frequencies, followed closely by the solid connection configuration.

Equipment Design↗

Phantom design method for high-field MRI human systems.

A phantom design method suitable for high-field MRI based on the RF field wave characteristics of sample and experimental validations at 7.0 T and 3.0 T are presented. The RF field distribution in a phantom with a given RF coil system is primarily determined by the sample size relative to the wavelength inside the sample, and the ratio between the displacement and conduction currents. Experimental results demonstrate that the MR image intensity patterns associated with wave behavior in human samples at a given field strength can be reproduced with a phantom at the same or different field strengths once the dimension and penetration constant are scaled by the corresponding wavelength in the sample medium.

Brain Mapping↗

Spatial resolution of numerical models of man and calculated specific absorption rate using the FDTD method: a study at 64 MHz in a magnetic resonance imaging coil.

PURPOSE: To examine how fine a model resolution is necessary for calculation of specific energy absorption rate (SAR) for the human head in regions as small as 1 g. MATERIALS AND METHODS: Here we perform a simple study comparing the maximum SAR averaged over any 1 cm(3) and SAR averaged over the entire head for several models of the same human head within the same radiofrequency coil, but with spatial resolutions varying from 8-100 Yee cells per cm(3). RESULTS: Over the range of model resolutions from 8-100 Yee cells per cm(3), there is only a 16% variation in maximum SAR in any 1 cm(3) of tissue in the head, and only a 7% variation in SAR averaged over the entire head. CONCLUSION: While it is always desirable to perform SAR calculations with the greatest possible accuracy, in calculations of the maximum SAR levels in any 1 cm(3) of tissue, spatial resolutions greater than 5 mm may not yield notably different results than those with a spatial resolution of 5 mm.

Head↗

Theoretical and experimental evaluation of detached endcaps for 3 T birdcage coils.

The use of detached endcaps for 3 T birdcage coils was investigated both theoretically and experimentally. Finite difference time domain analysis, along with workbench and MRI techniques, were used to map the radiofrequency (RF) B(1) distribution along the coil axis with and without an endcap. Without an endcap the measured B(1) value at the service end of the birdcage was only 45% of the value at the coil's center. This was improved to 85% with a detached endcap of maximum achievable diameter (375 mm), positioned 4 mm from the RF shield. The B(1) field distribution on the patient side of the coil was unaffected by the presence of the endcap. The dependence of the B(1) distribution as a function of endcap diameter was also investigated. Surprisingly, simulations and experiments show that there is an optimum ratio of endcap-to-birdcage coil diameter (approximately 1.08) that gives the best B(1) homogeneity. In the human head the optimized endcap, positioned 16 mm from the RF shield, improves the MRI signal amplitude from 55% to 85% of maximum toward the service end. This novel endcap design is easy to implement with existing birdcage coils, and could prove useful when flexibility in access to the RF coil is required.

Equipment Design↗

Analysis of wave behavior in lossy dielectric samples at high field.

Radiofrequency (RF) field wave behavior and associated nonuniform image intensity at high magnetic field strengths are examined experimentally and numerically. The RF field produced by a 10-cm-diameter surface coil at 300 MHz is evaluated in a 16-cm-diameter spherical phantom with variable salinity, and in the human head. Temporal progression of the RF field indicates that the standing wave and associated dielectric resonance occurring in a pure water phantom near 300 MHz is greatly dampened in the human head due to the strong decay of the electromagnetic wave. The characteristic image intensity distribution in the human head is the result of spatial phase distribution and amplitude modulation by the interference of the RF traveling waves determined by a given sample-coil configuration. The numerical calculation method is validated with experimental results. The general behavior of the RF field with respect to the average brain electrical properties in a frequency range of 42-350 MHz is also analyzed.

Electromagnetic Phenomena↗

Polarization of the RF field in a human head at high field: a study with a quadrature surface coil at 7.0 T.

The RF field intensity distribution in the human brain becomes inhomogeneous due to wave behavior at high field. This is further complicated by the spatial distribution of RF field polarization that must be considered to predict image intensity distribution. An additional layer of complexity is involved when a quadrature coil is used for transmission and reception. To study such complicated RF field behavior, a computer modeling method was employed to investigate the RF field of a quadrature surface coil at 300 MHz. Theoretical and experimental results for a phantom and the human head at 7.0 T are presented. The results are theoretically important and practically useful for high-field quadrature coil design and application.

Brain↗

Genetic requirements for the episomal maintenance of oncogenic herpesvirus genomes.

Herpesviruses are large double-stranded DNA viruses that are characterized by lifelong latency. Epstein-Barr virus (EBV), the recently discovered Kaposi's sarcoma associated herpesvirus (KSHV), also referred to as human herpesvirus-8 (HHV-8), and the simian Herpesvirus saimiri (HVS) are associated with malignant lymphoproliferative diseases. These viruses establish latent infection in lymphoid cells. During latency only a few viral genes are expressed and the viral genome persists as a multicopy circular episome. The episome contains repetitive sequences that serve as multiple cooperative binding sites for the viral DNA binding proteins Epstein-Barr virus nuclear antigen 1 (EBNA-1) of EBV and latency-associated nuclear antigen (LANA1) of KSHV and HVS, which are expressed during latency. The oligomerized proteins associate with the viral genome and tether it to host chromosomes, assuring continual lifelong persistence of the virus.

Antigens, Viral↗