MR thermometry characterization of a hyperthermia ultrasound array designed using the k-space computational method.

Abstract

BACKGROUND: Ultrasound induced hyperthermia is a useful adjuvant to radiation therapy in the treatment of prostate cancer. A uniform thermal dose (43 degrees C for 30 minutes) is required within the targeted cancerous volume for effective therapy. This requires specific ultrasound phased array design and appropriate thermometry method. Inhomogeneous, acoustical, three-dimensional (3D) prostate models and economical computational methods provide necessary tools to predict the appropriate shape of hyperthermia phased arrays for better focusing. This research utilizes the k-space computational method and a 3D human prostate model to design an intracavitary ultrasound probe for hyperthermia treatment of prostate cancer. Evaluation of the probe includes ex vivo and in vivo controlled hyperthermia experiments using the noninvasive magnetic resonance imaging (MRI) thermometry. METHODS: A 3D acoustical prostate model was created using photographic data from the Visible Human Project. The k-space computational method was used on this coarse grid and inhomogeneous tissue model to simulate the steady state pressure wavefield of the designed phased array using the linear acoustic wave equation. To ensure the uniformity and spread of the pressure in the length of the array, and the focusing capability in the width of the array, the equally-sized elements of the 4 x 20 elements phased array were 1 x 14 mm. A probe was constructed according to the design in simulation using lead zerconate titanate (PZT-8) ceramic and a Delrin plastic housing. Noninvasive MRI thermometry and a switching feedback controller were used to accomplish ex vivo and in vivo hyperthermia evaluations of the probe. RESULTS: Both exposimetry and k-space simulation results demonstrated acceptable agreement within 9%. With a desired temperature plateau of 43.0 degrees C, ex vivo and in vivo controlled hyperthermia experiments showed that the MRI temperature at the steady state was 42.9 +/- 0.38 degrees C and 43.1 +/- 0.80 degrees C, respectively, for 20 minutes of heating. CONCLUSION: Unlike conventional computational methods, the k-space method provides a powerful tool to predict pressure wavefield in large scale, 3D, inhomogeneous and coarse grid tissue models. Noninvasive MRI thermometry supports the efficacy of this probe and the feedback controller in an in vivo hyperthermia treatment of canine prostate.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Osama M Al-Bataineh, Christopher M Collins, Eun-Joo Park, Hotaik Lee, Nadine Barrie Smith. 2006-10-25. MR thermometry characterization of a hyperthermia ultrasound array designed using the k-space computational method.. https://doi.org/10.1186/1475-925x-5-56

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Effects of gamma irradiation on mechanical properties of defatted trabecular bone allografts assessed by speed-of-sound measurement.

New sterilization methods for human bone allografts may lead to alterations in bone mechanical properties, which strongly influence short- and medium-term outcomes. In many sterilization procedures, bone allografts are subjected to gamma irradiation, usually with 25 KGy, after treatment and packaging. We used speed-of-sound (SOS) measurements to evaluate the effects of gamma irradiation on bone. All bone specimens were subjected to the same microbial inactivation procedure. They were then separated into three groups, of which one was treated and not irradiated and two were exposed to 10 and 25 KGy of gamma radiation, respectively. SOS was measured using high- and low-frequency ultrasound beams in each orthogonal direction. SOS and Young modulus were altered significantly in the three groups, compared to native untreated bone. Exposure to 10 or 25 KGy had no noticeable effect on the study variables. The impact of irradiation was small compared to the effects of physical or chemical defatting. Reducing the radiation dose used in everyday practice failed to improve graft mechanical properties in this study.

Acoustics↗

Objective and subjective evaluation of the acoustic comfort in classrooms.

The acoustic comfort of classrooms in a Brazilian public school has been evaluated through interviews with 62 teachers and 464 pupils, measurements of background noise, reverberation time, and sound insulation. Acoustic measurements have revealed the poor acoustic quality of the classrooms. Results have shown that teachers and pupils consider the noise generated and the voice of the teacher in neighboring classrooms as the main sources of annoyance inside the classroom. Acoustic simulations resulted in the suggestion of placement of perforated plywood on the ceiling, for reduction in reverberation time and increase in the acoustic comfort of the classrooms.

Acoustics↗

Analysis of effective radiating area, power, intensity, and field characteristics of ultrasound transducers.

OBJECTIVE: To characterize the ultrasound fields produced by a cohort of transducers from a single manufacturer via hydrophone and Schlieren technology. DESIGN: Descriptive study. SETTING: Measurement laboratory. PARTICIPANTS: Seven same-model ultrasound transducers from a single manufacturer. INTERVENTIONS: Not applicable. MAIN OUTCOME MEASURES: Effective radiating area (ERA), total power, spatial average intensity (SAI), beam nonuniformity ratio (BNR), and Schlieren beam widths at 1.0 and 3.3 MHz. RESULTS: Values for ERA (1.0 MHz range, 3.62-4.38 cm(2); 3.3 MHz range, 3.74-4.76 cm(2)), total power (1.0 MHz range, 5.0-5.6 W; 3.3 MHz range, 4.7-5.7 W), SAI (1.0 MHz range, 1.2-1.4 W/cm(2); 3.3 MHz range, 1.0-1.5 W/cm(2)), and BNR (1.0 MHz range, 2.79-5.85; 3.3 MHz range, 2.51-4.56) fell within manufacturer's specifications and U.S. Food and Drug Administration (FDA) regulations. Schlieren analysis showed significantly larger beam widths at 3.3 MHz compared with 1.0 MHz and a large degree of variability in the ultrasound fields generated by the different transducers. There were no significant correlations between beam widths and ERA values. CONCLUSIONS: ERA and total power values in a test cohort exist within a range that met FDA regulations. Individual variability in ERA and total power resulted in 50% variability in SAI. This variability may help explain previous reports of heating differences between transducers.

Acoustics↗