Computer-controlled positive displacement pump for physiological flow simulation.

A computer-controlled pump for use both in the study of vascular haemodynamics and in the calibration of clinical devices which measure blood flow is designed. The novel design of this pump incorporates two rack-mounted pistons, driven into opposing cylinders by a micro-stepping motor. This approach allows the production of nearly uninterrupted steady flow, as well as a variety of pulsatile waveforms, including waveforms with reverse flow. The capabilities of this pump to produce steady flow from 0.1 to 60 ml s-1, as well as sinusoidal flow and physiological flow, such as that found in the common femoral and common carotid arteries are demonstrated. Cycle-to-cycle reproducibility is very good, with an average variation of 0.1 ml s-1 over thousands of cycles.

Arteries

Computer-controlled flow simulator for MR flow studies.

A novel computer-controlled flow simulator for use in magnetic resonance (MR) flow experiments was evaluated. The accuracy in constant-flow mode was better than 1%. The accuracy in pulsatile-flow mode was found to be dependent on the interconnecting tubing. The short-term and long-term reproducibilities of pulsatile waveforms were less than or equal to 0.4 mL/sec (1 standard deviation). Increased response times due to the lengths of tubing required in MR flow experiments were surmounted by using a modified tubing configuration and precompensated waveforms. Piston reversal was found not to cause major difficulties in MR flow experiments.