PubMed Health⌕ Search

Biomedical subjects

M Hexamer

Publications and source records attributed to M Hexamer.

At least 19 recordsLinked to original sources

Robust and self-tuning blood flow control during extracorporeal circulation in the presence of system parameter uncertainties.

Three different discrete controllers were designed and tuned to be used in conjunction with a rotary blood pump during cardiopulmonary heart-lung support. The controllers were designed to operate in both steady and pulsatile modes. The system and methods were tested in a circulatory haemodynamic simulator. To guarantee stable control of the non-linear circulatory system in the presence of patient parameter uncertainties, a proportional plus integral (PI) and an H infinity controller were robustly tuned, using a non-linear time-varying model. (H infinity refers to the Hardy space, the set of bounded functions, analytic in the right half plane. The H infinity controller is the solution to the H infinity norm optimisation problem.) A self-tuning general predictive controller (GPC), together with an adaptive Kalman filter (KF) estimator, was compared with the two robustly tuned controllers. The closed-loop blood flow control circuit was set up in simulation routines first. The blood flow controllers were validated in a circulatory hydrodynamic simulator (MOCK) combined with a rotary blood pump. Parameters of the system simulator were changed continuously, and the controllers were tested over a wide range of different operating points. Disturbances in the form of discontinuous additive parameter uncertainties were applied. The closed-loop systems remained robustly stable. The robustly tuned H infinity controller showed the best control performance, in contrast to the GPC controller, which was near instability in regions of strongly varying non-linear system gain. Compared with the H infinity controller, the PI controller showed slightly worse behaviour, but the closed-loop response was acceptable, even in regions of strongly varying non-linear system gain and during pulsatile perfusion. The rotary blood pump could provide stationary and pulsatile perfusion under control conditions. Controlled variables were hereby mean blood flow, pulsatility index and heart rate. All three controllers were developed for an arterial mean flow of 0-6 l min(-1) and a heart rate of up to 70 beats per minute. Pulsatile closed-loop perfusion could provide up to 30 mmHg pressure variation in the simulated ascending aorta at a mean flow of 3 l min(-1).

Aorta↗

Rate-responsive pacing using the atrio-ventricular conduction time: design and test of a new algorithm.

The cardio-circulatory system of chronotropic incompetent patients is unable to adapt heart rate adequately to the level of strain. A common therapy is the implantation of a pacemaker that stimulates the heart at a rate derived from a strain related sensor signal. The paper describes a new pacemaker concept that uses a well-defined time interval in the electrogram as a sensor parameter, the atrioventricular conduction time (AVCT). Identification experiments with patients delivered the stationary and dynamic behaviour of AVCT subject to variations in the exercise rate and the pacing frequency. Furthermore, it was demonstrated that AVCT is perturbed by respiratory activity. The new rate-responsive algorithm, which uses the internal model control principle, explicitly takes into account the closed-loop nature of the underlying system. The major design objectives were: extended range of the individual heart rate; effective attenuation of the respiratory-related disturbance; and stability. Seven patients undergoing incremental exercise were paced with this AVCT-based algorithm. The experiments confirmed the suitability of this concept with respect to the design goals. The patients' peak intrinsic heart rate at exercise was 100 +/- 20min(-1). When paced with the AVCT algorithm, the paced heart rate was 126 +/- 12min(-1). The increase was significant (26 +/- 13min(-1); 15 to 53 min(-1)), which clearly demonstrated the potential of this concept to restore chronotropic competence.

Aged↗

Mechanical and optical characteristics of a new fiber optical system used for cardiac contraction measurement.

In order to obtain a better physiological performance and a closer restoration of the regular rhythm of failing hearts, a new fiber optical sensor system for the measurement of cardiac contraction has been developed. It consists of an opto-electrical unit and a sensing fiber which has to be positioned in the heart. The objective of this new fiber optic sensor system is to use the inotropic information to adjust a stimulation algorithm in single or multichamber pacing or to detect arrhythmia in insufficient heart function. In this study, the mechanical and optical characteristics of different fibers are investigated. The relationship between the attenuation (with an achieved numerical maximum of 0.3 dB), the bending diameter and the angle of bending is determined in a range of 20-160 mm. The most suitable fiber for the application in cardiological problems is determined (WT8 fiber), for which the sensitivity is analyzed. Additionally, power spectra are calculated from WT8 fiber signals obtained from pig hearts, working under physiological conditions. The maximal frequency response was 23 Hz. It is concluded that the fiber optical measurement of cardiac contraction is not only feasible and reproducible, but the WT8 fiber also shows optimal behavior in the range of parameters occurring in the heart chambers. Nevertheless, in order to restrict the measured signal reliably to bending processes within the chambers only, it is concluded that a special combined fiber has to be constructed with a high sensitivity only at its terminal section within the heart.

Algorithms↗

The dromotropic pacemaker: system analysis and design considerations.

Patients suffering from chronotropic incompetence are generally treated with a rate-responsive pacemaker that stimulates the heart at a rate derived from a strain related sensor signal. The pacemaker concept described here uses a well-defined time interval in the electrogram as sensor parameter (AVCT: atrio-ventricular conduction time). AVCT is directly controlled by the autonomic nervous system. The design of the new algorithm was based on a thorough experimental analysis of AVCT subject to variations of the exercise rate and the pacing frequency. There it was demonstrated that AVCT is disturbed by the respiratory activity. The new rate-responsive algorithm which uses the internal model control principle explicitly takes into account the closed-loop nature of the underlying system. The major design objectives were: a) extended range of the individual heart rate, b) effective attenuation of the respiratory related disturbance and c) dynamic stability. Seven patients undergoing an incremental exercise test were paced with the new AVCT-based algorithm. When paced with this algorithm the paced heart rate was 126 +/- 12 min(-1) whereas the peak intrinsic heart rate was only 100 +/- 20 min(-1). The increase which was significant (26 +/- 13 min(-1); 15.53 min(-1)) clearly demonstrated the potential of this concept to restore chronotropic competence. A reanalysis of the experiments was undertaken in order to facilitate the individual parameterization in clinical practice. It could be demonstrated that a rather simple screening test is sufficient to gain the knowledge necessary for the individual parameterisation.

Algorithms↗

Morphological detection algorithms for the automatic implantable cardioverter/defibrillator (AICD).

To prevent sudden cardiac death of patients who are at risk from long standing tachyarrhythmia the implantable cardioverter defibrillator (ICD) is the first choice therapy. ICDs use a range of electrostimuli up to defibrillation, which is a non synchronous high energy shock, whereas cardioversion is synchronous with the ECG. In order to know when and how to react, a detection algorithm, which analyses an intracardial electrocardiogram (ECG) and classifies the heart rhythm, is implemented in every ICD. All detection algorithms use the heart rate to classify the different heart rhythms roughly. If a tachycardia is detected, it is important to discriminate between a ventricular tachycardia, which is life threatening and a supraventricular tachycardia, which is much less threatening. To be able to make this distinction the detection algorithms analyse the behaviour of the heart cycle intervals, the ECG-morphology or in addition to the ventricular ECG, an atrial ECG. In this paper morphological algorithms will be evaluated and newly developed algorithms will be presented. Recent algorithms use the mathematical wavelet theory. The evaluation shows that these get better results than all but one of the simpler classical morphological algorithms. A new wavelet based algorithm, developed by the authors, exhibits the best detection results.

Algorithms↗

Automatic control of the extra-corporal bypass: system analysis, modelling and evaluation of different control modes.

Automatic control of the blood gas parameters during extracorporeal circulation has the potential to improve the quality of this procedure and to relieve the personnel from a time consuming task. This paper describes a model of the underlying system for a standard clinical set-up and pinpoints the major difficulties which are the variations of the process gains and the blood- and gas-flow dependent dead times and time constants. Scheduled PI-controllers both for the arterial oxygen as well as for the carbon dioxide partial pressure were designed. Scheduling was based on the blood flow rate. These controllers were tested in a simulation environment. The control systems remained stable under all tested operating condition, but if the blood flow rate was changed abruptly rather large load errors occurred. The performance was improved markedly by adding a feed-forward control path which directly influences the actuating signals based on the actual blood flow rate and the hemoglobin contents, variables which are measured anyway. The major conclusion of this study is to use such direct feed-forward compensation even if more sophisticated control algorithms are used.

Algorithms↗

Automatic control of isolated organ perfusion.

If provided with adequate physiological conditions explanted hearts may continue to operate in the regular beating mode. This property offers the opportunity to substitute extensive animal experiments by investigations on the isolated organs. At the same time it enables the development of an alternative method for the transport of donor hearts for transplantation. Experimental setups for both applications are described in this paper. The focus is laid on the optimisation of handling and effectiveness of these setups by means of automatic control. The method for investigating the controlled system ("plant") for blood gas exchange, and especially the determination of the transfer function for the partial pressure of oxygen are described. The plant was triggered with stepwise changes on the gas side at several operating points. A first order lag element with time delay was chosen as approximation of every transfer function. The parameter 'time delay', 'gain of the plant' and 'time constant' were analysed as to the dependency on blood flow rate, gas flow rate and partial pressure of oxygen at the blood outlet of the oxygenator. As a result an equation was found to calculate the time delay from gas flow rate and blood flow rate. Correlation of gain and time constant with parameters of the plant were obtained, too. The data is used for the design of a controller, adapting to the different operating points of the plant.

Algorithms↗

Investigation of cardiac and cardio-therapeutical phenomena using a pulsatile circulatory model.

For many years computer models of the cardiovascular system have existed with which the complex physiological interrelations of regulation processes may be studied. However, one major disadvantage of these models is the simplified modeling of the heart kinematics whereby the short term regulation effects of the heart itself on varying physical strains may only be simulated rudimentarily. The developed pulsatile model emphasizes the detailed description of the systolic wall tension progression subject to the Hill/Frank-Starling mechanisms and the simulation of the cardiac valves. Moreover, to be able to describe long term effects the pulsatile model was coupled to the established cardio vascular model of Guyton. Using the coupled system the influences of impaired valve function and ventricular relaxation disorders on the cardiac output as well as the need for an implementation of a physical strain dependent av-delay in existing pacemaker types were investigated. Both the dynamics and the steady-state values of a subset of variables (pressure, volume, flow) were successfully validated by comparing them to published experimental and clinical recordings.

Biomechanical Phenomena↗

New sensor based on fibre optics for measurement of heart movement.

Innovative fibre-optic sensor technology for measuring the movement of the myocardial walls, and from this the heart chamber volumes, was developed. An optical fibre, with a mirror at its end, is inserted into a catheter located in the heart. An opto-electrical control unit positioned outside the heart contains both the light source and the signal receiver. It generates and couples the light into the fibre and transforms and analyses the reflected signal. With such a system, the movement of the cardiac wall can be continuously measured during each cycle, because the fibre moves synchronously with the heart, and this movement bends the fibre, changing the optical attenuation. Experiments where the fibres were wound around metal cylinders of different diameters revealed a maximum sensitivity of 4% mm(-1), diameter. The noise signal corresponded to about 1% of the diameter. First tests in a working pig heart showed a high correspondence of the fibre signal with cardiac parameters. Although these tests are promising, further long-term, extensive experiments in preclinical test devices, and later in clinical tests, must be carried out before the new sensor is used in clinical practice. The fibre-optic technique could be used in monitoring devices, assist devices, pacemaker systems or cardioverter defibrillators.

Animals↗

[Biventricular defibrillation using transvenous electrodes].

This study investigated the feasibility of transvenous biventricular defibrillation using an electrode in a left ventricular vein. The standard lead configuration with a coil in the right ventricle (RV) and a coil in the superior vena cava (SVC) was compared with an additional unipolar coil in an accessible epicardial vein. Only biphasic shocks were used with different shocking modes between the coils in the RV, LV, SVC and the ICD-generator (CAN). Shocks were applied starting with 30 J, decreasing till the DFT was reached. As a result there is a lower DFT when defibrillation is performed including the left ventricular electrode. The impedance did not show a significant increase after more then 20 consecutive shocks. It is a feasible and workable application that might help to reduce the energy demand and increase the safety of such a system.

Animals↗

[Rate-responsive pacemaker concept: parametric system identification].

The focus of this paper is the design of a rate-responsive pacemaker which uses the atrio-ventricular conduction time (AVCT) as an exertion related sensor signal. AVCT can be easily obtained from an intracardial electrogram. During atrial pacing AVCT shortens with exercise but increases with the stimulation rate. Hence, an AVCT based pacemaker always establishes a closed-loop system. General design rules for an AVCT pacemaker have been developed from our experimental results and a system-theoretical treatment. Topics addressed were the controller gain, uncertainties concerning the dynamics of the AVCT, the attenuation of disturbances, the closed-loop bandwidth and stability.

Algorithms↗

[Blood gas regulation in isolated beating pig hearts].

Continuous perfusion with oxygenated blood is required to supply isolated beating pig hearts. To meet physiological conditions the blood gas parameters pH, pO2 and pCO2 should correspond with values usually found in pigs, respectively. Controlling these parameters manually is a very time consuming task and is therefore done best by a control system consisting of gas blender, oxygenator and blood gas sensors. As this control system should be used with an transportable perfusion apparatus, this paper describes an approach where the gassing is realized only with air and oxygen. In this approach pO2 and pH are measured on blood side so that the control system can change oxygen concentration and gas flow adequately on the gas side.

Acid-Base Equilibrium↗

Assessment of the chronotropic response at the anaerobic threshold: an objective measure of chronotropic function.

The evaluation of the heart rate response to exercise is important for the diagnosis of chronotropic incompetence and the assessment of a rate responsive algorithm of sensor-controlled pacemakers. The aim of the present study was to examine a classification of the chronotropic response at an individually moderate exercise level. Sixteen pacemaker patients (patient group, age 62.9 +/- 7.6 years) with sick sinus syndrome and 15 age-matched healthy subjects (control group, age 57.6 +/- 9.4 years) underwent a maximum cardiopulmonary exercise test on a treadmill after a protocol with individually selected incremental steps. To analyze the patients' intrinsic heart rate response, the rate responsive mode of the pacemaker was switched off. Chronotropic incompetence was diagnosed in eight patients whose maximal heart rate was < 80% of the age-predicted heart rate. The heart rate at the anaerobic threshold was significantly lower in the chronotropically incompetent subgroup than in the chronotropically competent patients and the healthy subjects (85.9 +/- 6.6 beats/min vs 100.3 +/- 9.9 beats/min and 112.9 +/- 11.7 beats/min, respectively). The chronotropic slope of the heart rate reserve as a function of the metabolic reserve was significantly higher in the control group than in the patient groups with either mild or severe chronotropic incompetence (0.94 +/- 0.17 vs 0.64 +/- 0.08 and 0.43 +/- 0.14, respectively). Furthermore, the chronotropically incompetent response could be divided into a linear type with and without a threshold, an exponential, and a logarithmic type. The anaerobic threshold was an objectively detectable breakpoint at an individually moderate exercise level that could be used for characterization of chronotropic function. At the anaerobic threshold, a physiological heart rate response was about 220--age--50 beats/min. A deviation of more than 10 beats/min below this physiological value characterized chronotropic incompetence.

Algorithms↗

Multi-loop control of liquid cooling garment systems.

In this study the effects of multi-loop control of liquid cooling garments (LCGs) under exercise heat stress conditions were investigated by experiments and theoretical analysis. A triple-loop LCG, by which the torso, arms and legs could be independently cooled, was used in the two series of experiments carried out in a hot environment (35 degrees C/40% RH). The experiment consisted of rest, exercise on an ergometer at 70 W and exercise at 110 W. In the first experiment, each water inlet temperature (TWI) was adjusted according to the local thermal sensation. In the second experiment, TWI for the torso including arms and TWI for the legs were regulated by a skin temperature controller with set-point adjustment via heart rate. The experiments showed that a multi-loop LCG is more effective than a single-loop LCG in providing thermal sensation and comfort adjusted to the requirements of the different parts of the body, and that a skin temperature controller could be applied to a multi-loop system. The theoretical analysis was carried out using a mathematical model of thermoregulation. The results showed that a strong cooling of the surface over the working muscles (legs) provided the greatest thermoregulatory advantage during low body exercise, because most of the heat generated within the working muscles can be removed directly by heat conduction to the skin. Optimization of a human/LCG system could be attained by an optimal configuration and control. However, an optimal configuration always depends on the application purpose of an LCG system.

Adult↗

Restoration of cardio-circulatory regulation by rate-adaptive pacemaker systems: the bioengineering view of a clinical problem.

In the past, the development of rate-adaptive (sensor-controlled) pacemaker systems seems to have been determined primarily by the availability, compatibility and other properties of the technical sensor. This paper, however, focuses on the system-physiological aspect in an attempt to answer the question to what extent physiological cardiovascular control is restored by the pacemaker system. This is a question which should be asked before attempting to design a sensor-controlled system and especially before designing multisensor systems with infinite combinations. Four categories are defined: direct bridging ("shunting"), open loop systems, closed systems using cardiorespiratory or metabolic coupling and those using cardiac signals. Further subdivisions are shown. From the bioengineering as well as from the physiological viewpoint a system should preferably not combine sensors from one and the same of these categories. At present direct bridging is available only for the atrioventricular (AV)-block, so that for sick-sinus-syndrome (SSS) patients feedback control via cardiac signals ("inotropic" pacemaker) comes nearest the goal without, however, ideally bridging the gap. Open-loop systems should no longer be developed as single-sensor systems. A well developed activity sensor, however, which quickly pinpoints the most prominent stressor of cardiovascular control is best suited to complement another sensory system achieving closed-loop control. New and promising concepts orientated toward direct bridging are the analysis of monophasic action potentials and the "dromotropic" concept, both of which seek direct correlation with the "chronotropic" information not available in SSS patients.

Adaptation, Physiological↗

Influence of the pacing rate on the atrioventricular conduction time during aerobic and anaerobic exercise: basic concepts for a dromotropically controlled rate responsive pacemaker.

The dromotropic pacemaker concept needs a rate responsive algorithm in which the pacing rate is controlled by the atrioventricular conduction time (AVCT). To develop basic concepts for such a rate responsive algorithm, the influence of the pacing rate on the AVCT was investigated. Seven patients (62 +/- 7.8 years) with sick sinus syndrome and intact atrioventricular conduction underwent two cardiopulmonary exercise tests (CPX) on a treadmill. According to the determination of the anaerobic threshold (AT) and the patients maximum capacity in the first incremental CPX the work rate for two exercise levels below and above the AT were chosen for the second constant workload CPX. The calculation of the optimal pacing rate (HRopt) was based on the oxygen uptake (VO2) during exercise after reaching steady-state conditions. According to the increase of the VO2 from 14.8 +/- 2.3 mL/min per kilogram during aerobic work (38.3 +/- 16.0 W) to 19.4 +/- 4.7 mL/min per kilogram during anaerobic work (80.6 +/- 32.3 W), the HRopt was calculated to be 98.6 +/- 6.9 beats/min and 116.4 +/- 4.7 beats/min. Starting from HRopt, the pacing rate was increased (overpacing) and decreased (underpacing) by about 5 beats/min every minute. At optimal pacing rate the AVCT decreased significantly from 233.0 +/- 30.5 ms during aerobic work and to 226.4 +/- 27.3 ms during anaerobic work (P < 0.05). Whereas overpacing induced a significant prolongation of the AVCT during aerobic work (4.17 +/- 1.78 ms per 10 beats/min) and anaerobic work (3.84 +/- 1.60 ms per 10 beats/min), underpacing yielded a significant shortening of the AVCT by about 4.49 +/- 2.64 ms per 10 beats/min during aerobic work and 4.75 +/- 1.87 ms per 10 beats/min during anaerobic work (P < 0.01). The slopes of the regression lines of the relationship between AVCT and pacing rate were not significantly. different. Based on the reciprocal relationship of heart rate (HR) and AVCT, basic concepts may be established for a dromotropic rate responsive algorithm.

Aged↗

Relationship between atrioventricular delay and oxygen consumption in patients with sick sinus syndrome: relevance to rate responsive pacing.

To develop a dromotropic-controlled rate adaptive algorithm for patients with sick sinus syndrome (SSS) and intact AV conduction, 14 pace-maker patients with SSS underwent cardiopulmonary exercise testing (CPX). During exercise, the pace-maker was programmed in an AAT mode without rate adaptation, whereby 3 patients developed supraventricular arrhythmia and 11 patients kept sinus rhythm. Chronotropic incompetence (CI) at heart rate (HR) < 95 beats/min at the anaerobic threshold (AT) was found in five patients. In patients with chronotropic competence (CC), the HR increase was significantly greater than in CI patients (rest: 73.2 +/- 12.6 vs. 64.2 +/- 4.0 beats/min;AT:101.2 +/- 6.2 vs. 82.0 +/- 5.1 beats/min;peak: 135.2 +/- 10.7 vs. 103.2 +/- 10.9 beats/min). There was no significant difference in the AVD between CC and CI patients (rest: 167.7 +/- 38.6 vs. 170.8 +/- 22.5 ms, AT: 156.2 +/- 30.7 vs. 163.6 +/- 21.6 ms, peak: 144.7 +/- 29.0 vs. 152.4 +/- 15.0 ms). The correlation coefficient between HR increase and VO2 was +1.0 and between AVD decrease and VO2 - 1.0 in both groups. An increase in pacing rate from 75 beats/min to 120 beats/min without exercise (overpacing) led to a prolongation of the AV interval of about 30.6 +/- 14.2 ms. Based on this closed loop control with negative feedback, a dromotropic rate adaptive algorithm for patients with SSS and intact AV conduction could be developed.

Aged↗