Control of liquid cooling garments: technical control of mean skin temperature and its adjustments to exercise.

This paper describes an automatic control concept for liquid cooling garments. The concept consists of an own controller for mean skin temperature whose setpoint is either fixed or adjusted to the metabolic heat production by means of the heart rate signal. The controller for mean skin temperature included both a proportional and an integral signal path (PI-type), the latter being able to eliminate any load error within the control loop. This means that the actual skin temperature will always match the given setpoint irrespective of the amount and the origin of the heat gain at the body shell. Experiments were carried out to test the operation of the skin temperature controller. There the setpoint was fixed while metabolic heat production was changing. After a transient period with deviations, the load error was always eliminated by the skin temperature controller. With this result one can also imagine the controllers ability to compensate changing heat gains from the environment. Despite this behaviour, the amount of heat removal was not high enough to prevent sweating and warm discomfort during all exercise levels. Therefore we draw the conclusion that, in addition, the setpoint of the skin temperature should be adjusted to the metabolic rate/heat production. A convenient physiological signal that reflects the current level of metabolic rate is the heart rate signal. After being filtered the heart rate signal was used during some experiments to change the setpoint of the skin temperature controller. The reason for this filtering (lowpass, time constant = 10 min) was, firstly, the necessity of attenuating the consequences of short-term psychological effects on the heart rate and secondly, the avoidance of vasoconstriction due to too fast changes of the exercise/heart rate induced cooling rate. In the following experiments it became clear that the adjustment of the skin temperature setpoint to the exercise level was an improvement as there was less sweating and the subjects felt more comfortable.

Automation↗

Control of liquid cooling garments: technical control of body heat storage.

This paper describes a concept of how liquid cooling garments (LCG) can be automatically controlled by the objective physiological state. The technically controlled parameter was mean body temperature which was calculated from the measured rectal and mean skin temperature. This was motivated by the fact that mean body temperature is the basis for estimating body heat storage, a commonly used measure of thermal strain. Here the setpoint of mean body temperature was the individual value taken in a preceeding resting period and it was the task of the technical controller to keep the actual value of mean body temperature as close as possible to the setpoint. The most important tuning parameters of the controller were the weighting coefficients for rectal and mean skin temperature in the calculation for mean body temperature. The ratio of these two coefficients determined the degree of compensation for any rectal temperature shift by changing mean skin temperature. Test experiments were carried out (n = 5) in which the controller was able to clamp mean body temperature to the setpoint thereby preventing heat storage. Although exercise rate (75 W) was the same, sweating and warm discomfort occurred in some cases due to the individual rectal temperature rise. Another source of discomfort were delays or paradoxical time courses of rectal temperature at the start or end of exercise which were responsible for a delayed onset of cooling or heating. To avoid these effects, the oxygen consumption signal, which is very fast and directly correlated to the exercise rate, was added to the control loop. Each increase of this parameter above its resting level lowered suit temperature. As heat storage should not be completely rejected by this new signal pathway, the controller for mean body temperature still remained active. The repetition of the experiments showed that the load error in the control loop was smaller and the comfort level in transient phases higher. For a further improvement of this concept it is recommended that the weighting coefficients be tuned to the individual requirements.

Adult↗

Control of liquid cooling garments: subjective versus technical control of thermal comfort.

Liquid cooled garments (LCG) are a powerful tool for alleviating heat strain during work in hot conditions. However, the potential advantage of an LCG depends on a more or less proper control of the cooling liquid's temperature. To gain more knowledge on this subject two experimental studies concerning manual control by the wearer were carried out. In both studies the subjects had to exercise in a warm environment. In the first study the subjects (n = 5) were asked to control the temperature directly. They used different strategies, ranging from gradual changes to oscillations of the suit temperature. This was accompanied by large differences of the chosen inlet temperature and the comfort level attained, although by repetition of the experiment it was possible to increase the subjective thermal comfort. In the second study a technical 'comfort' controller changed the inlet temperature of the LCG in the appropriate direction whenever the personal assessment of thermal sensation differed from the neutral state. The subjects (n = 6) had no information about the control mode. The experiments showed that it was possible to maintain a comfortably neutral thermal sensation throughout nearly the whole experiment. This result, however, was partly in contrast to the objective thermal state of the subjects. Two of them had problems with their heat balance, expressed by the fact that their rectal temperature did not reach a steady state. The results of both studies indicate that manual or subjective control is not optimal and should therefore be replaced by an objective control of the thermal state.

Body Temperature↗

Computer supported analysis of cardiovascular parameters by impedance cardiography and plethysmography.

A fully automated system for computer supported analysis of cardiovascular parameters on the basis of impedance technique, which is confined to the analysis of relative changes of stroke volume and cardiac output, has been developed and tested. The evaluation showed good agreement between automatic and manual calculation. As the whole computational procedure is done by software, which is written in a common language (BASIC), a maximum of flexibility and portability is guaranteed. Estimation of relative changes of stroke volume and cardiac output as well as peripheral blood flow is possible without any manual interaction. Extension from physiological measurement of healthy individuals to clinical application has been started.

Cardiac Output↗