Digital radiography of the chest: design features and considerations for a prototype unit.

The general features of a prototype digital chest unit are described along with the rationale for the choice of design factors employed. It is shown that the scanning-slit, linear-detector-array approach employed can, with available x-ray tube technology, achieve a spatial resolution of 1 cy/mm and detector radiation levels comparable with those obtained with conventional screen-film systems. Also discussed are the unit's exposure latitude and its ability virtually to eliminate scatter.

Computers

Measured performance characteristics of a solid-state linear detector array.

Experimental studies of a solid-state linear detector array developed for a prototype scanning slit digital chest radiographic unit have been completed. The detector consists of a strip of scintillating material, optically coupled to a linear silicon photodiode array. Measured performance characteristics of the detector, such as sensitivity, modulation transfer function, and detective quantum efficiency, are presented for several different scintillators. Results indicate that direct x-ray absorption events in the silicon photodiode can degrade detective quantum efficiency. Results also indicate that the inexpensive preamplifier circuits used in the digital chest prototype contribute negligible noise at diagnostic x-ray photon fluence rates.