Measured performance characteristics of a solid-state linear detector array.
Abstract
Experimental studies of a solid-state linear detector array developed for a prototype scanning slit digital chest radiographic unit have been completed. The detector consists of a strip of scintillating material, optically coupled to a linear silicon photodiode array. Measured performance characteristics of the detector, such as sensitivity, modulation transfer function, and detective quantum efficiency, are presented for several different scintillators. Results indicate that direct x-ray absorption events in the silicon photodiode can degrade detective quantum efficiency. Results also indicate that the inexpensive preamplifier circuits used in the digital chest prototype contribute negligible noise at diagnostic x-ray photon fluence rates.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R A Sones, D R Morgan, M M Tesic, R A Mattson. Measured performance characteristics of a solid-state linear detector array.. https://doi.org/10.1118/1.595770