Effect of anterior capsule polishing on fibrotic capsule opacification: three-year results.

PURPOSE: To evaluate the long-term effect of anterior capsule polishing on anterior capsule opacification (ACO) and peripheral fibrotic posterior capsule opacification (PCO). SETTING: Department of Ophthalmology, Medical University of Vienna, Vienna, Austria. METHODS: This randomized double-blind study comprised 104 eyes of 52 patients with bilateral age-related cataract. All patients received round-edged intraocular lenses (IOLs); 26 received an SI-40 IOL (Advanced Medical Optics Inc.) in both eyes, and 26 received a Silens6 IOL (Domilens) in both eyes. Both IOLs consist of different silicone material and have different haptic angulation. The SI-40 IOL has 13.0 mm open-loop poly(methyl methacrylate) (PMMA) haptics angulated by 10 degrees. The Silens6 IOL has 12.5 mm open-loop PMMA haptics with no angulation. In 1 eye, the anterior capsule was extensively polished. The anterior capsule was left unpolished in the contralateral eye, which acted as a control. Digital slitlamp photographs of the ACO and fibrotic PCO were taken with a standardized technique for 3 years postoperatively. The intensity of ACO was measured objectively (score 0% to 100%) using Adobe Photoshop software. Fibrotic PCO was graded subjectively (score 0 to 4). RESULTS: The mean ACO was 17% in the polished eyes and 26% in the control eyes (P = .0001). The mean fibrotic PCO score was 0.5 and 1.0, respectively (P = .0007). The mean ACO was 15% in the polished SI-40 eyes and 26% in the control SI-40 eyes (P = .01). It was 19% in the polished Silens6 eyes and 26% in the control Silens6 eyes (P = .003). The mean fibrotic PCO score was 0.4 in the polished SI-40 eyes and 1.1 in the control SI-40 eyes (P = .0006). It was 0.6 in the polished Silens6 eyes and 0.9 in the control Silens6 eyes (P = .08). CONCLUSIONS: Three years after surgery, eyes in which the anterior capsule was extensively polished had less ACO and fibrotic PCO with both round-edged silicone IOLs. In eyes with Silens6 IOLs, however, the reduction in fibrotic PCO was not significant.

Aged↗

Laserinterferometric assessment of pilocarpine-induced movement of an accommodating intraocular lens: a randomized trial.

PURPOSE: To measure the axial movement of an accommodating intraocular lens (IOL) induced by ciliary muscle contraction after application of pilocarpine. DESIGN: Randomized, controlled, patient- and examiner-masked trial with intrapatient comparison. PARTICIPANTS AND CONTROLS: One hundred ten eyes of 55 patients with age-related bilateral cataract. METHODS: This study was divided into 3 parts. In the first, the accommodating IOL (1CU) was compared with a 3-piece open-loop acrylic IOL that served as the control. In the second, to assess the effect of capsule fibrosis on the potential accommodating performance of the accommodating IOL, extensive polishing of the anterior capsule with a slit cannula was compared with standard surgery. In the third, the effect of a posterior capsulorhexis was compared with that of standard surgery. Anterior chamber depth (ACD) was assessed with partial coherence interferometry, measured before and after topical application of pilocarpine 2%, and near visual acuity (VA) was evaluated 3 months after surgery. MAIN OUTCOME MEASURE: Pilocarpine-induced change in ACD. RESULTS: The accommodating IOL showed a forward movement under pilocarpine with a median amplitude of movement of -314 microm (95% confidence interval [CI]: -148 to -592), compared with the backward movement of 63 microm (95% CI: 161 to -41) for the open-loop control IOL (P = 0.001). Capsule polishing and a posterior capsulorhexis had no effect on IOL movement with the accommodating IOL. The median near VA with distance correction was 20/60. CONCLUSION: Pilocarpine induced a small but significant forward movement of the accommodating IOL. However, the amount of movement was calculated to result in a refractive change of <0.5 diopters (D) in most patients, reaching 1 D or slightly more in only single cases, with a large variability of movement. Neither polishing of the capsule bag nor a posterior capsulorhexis could enhance the accommodative ability.

Accommodation, Ocular↗

Comparison of posterior capsule opacification between the 1-piece and 3-piece Acrysof intraocular lenses: two-year results of a randomized trial.

PURPOSE: To compare intensity of posterior capsule opacification (PCO) between the 1-piece and 3-piece haptic designs of an open-loop hydrophobic acrylic intraocular lens (IOL). DESIGN: A randomized, patient- and examiner-masked clinical trial with intraindividual comparison. PARTICIPANTS: Fifty-two patients with bilateral age-related cataract (104 eyes). METHODS: Each patient had cataract surgery in both eyes and received a 1-piece Acrysof IOL in one eye and a 3-piece Acrysof IOL in the fellow eye. Follow-up examinations were at 1 week, 1 month, 6 months, 1 year, and 2 years. Patients were examined at the slit lamp, visual acuity (VA) was determined, and standardized high-resolution digital retroillumination images of the posterior capsule were taken. The intensity of PCO was assessed subjectively at the slit lamp and objectively using automated image analysis software. MAIN OUTCOME MEASURE: Posterior capsule opacification score (scale, 0-10). RESULTS: There was no significant difference between IOL styles in best-corrected VA, rhexis/IOL overlap, capsular folds, and amount of anterior capsule opacification during the follow-up period. One year postoperatively, the amount of regeneratory PCO was higher for the 1-piece Acrysof eyes (image analysis software score: 1.3) than for the 3-piece Acrysof eyes (score: 0.9; P = 0.002). However, 2 years postoperatively, there was no significant difference between the 2 IOL styles (1-piece: 1.5; 3-piece: 1.3; P = 0.3). Neodymium:yttrium-aluminum-garnet capsulotomy was not performed in the study. CONCLUSION: One year postoperatively, the 1-piece Acrysof showed slightly more regeneratory PCO than the 3-piece Acrysof. However, 2 years postoperatively, the barrier effect of the 1-piece design was comparable to that of the 3-piece haptic design, with low PCO intensity.

Acrylic Resins↗

Effect of haptic design on change in axial lens position after cataract surgery.

PURPOSE: To compare the postoperative changes in anterior chamber depth (ACD) between single-piece and multipiece AcrySof intraocular lenses (IOLs) (Alcon Laboratories, Inc.) and their effect on the postoperative refractive shift. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This prospective randomized bilateral study with intraindividual comparison comprised 104 eyes of 52 consecutive patients with age-related cataract. All patients had standardized cataract surgery with a temporal self-sealing incision, phacoemulsification, and IOL implantation in the capsular bag. The first 30 patients received a single-piece, 5.5 mm optic IOL (AcrySof SA30) in 1 eye and a multipiece IOL (AcrySof MA30) in the other eye. Another 22 patients received the corresponding 6.0 mm optic models (AcrySof SA60 and AcrySof MA60, respectively). Anterior chamber depth measurements and evaluation of the change in ACD during the first 6 postoperative months were performed using partial coherence laser interferometry. The amount of change in capsulorhexis size during the first 6 postoperative months was assessed using standardized retroillumination photographs. RESULTS: Single-piece IOLs shifted significantly less postoperatively than multipiece IOLs. The multipiece IOLs shifted forward, especially from 1 day to 1 month. The change in the capsulorhexis area was similar between IOL types. CONCLUSIONS: Single-piece AcrySof IOLs shifted minimally after the first postoperative day; therefore, the prescription for spectacles in eyes with this IOL can be obtained soon after surgery.

Anterior Chamber↗

Effect of optic edge design and haptic angulation on postoperative intraocular lens position change.

PURPOSE: To assess the effect of optic edge design and optic-haptic angulation of open-loop intraocular lenses (IOLs) on postoperative axial movement and the final position of the optic by measuring the anterior chamber depth (ACD) during the first postoperative year using partial coherence interferometry (PCI). SETTING: Department of Ophthalmology, Vienna General Hospital, Institute of Medical Physics, University of Vienna, Vienna, Austria. METHODS: In study 1, a 3-piece silicone IOL with nonangulated modified C-loop haptics (MicroSil, Dr. Schmidt) was implanted in 78 eyes of 39 patients; patients were randomized to receive a round-edged optic IOL in 1 eye and a sharp-edged optic IOL in the other eye. The ACD was measured by PCI 1 day, 1 week, 3 months, and 1 year after surgery. In study 2, a foldable, 3-piece acrylic IOL with modified 10-degree angulated J-loop haptics (AcrySof MA60BM, Alcon) was implanted in 32 eyes of 32 patients. The ACD was measured by PCI 1 day, 1 week, and 3 months after surgery. RESULTS: In eyes with a nonangulated silicone IOL, there was a significant postoperative change in ACD with both sharp-edged and round-edged designs (P<.01). There was forward movement of both IOL designs in the first week, with no significant difference between the 2 models. From 1 week to 3 months, there was backward movement of IOLs of both designs, with the sharp-edged IOL moving a significantly greater amount (P<.001). From 3 months to 1 year, IOLs with both optic edge designs moved slightly backward. Sixty-six percent of angulated IOLs showed continuous but variable forward movement and 34%, backward movement. CONCLUSIONS: Optic edge design influenced postoperative axial optic movement and thus had an impact on the development of postoperative refraction (refractive shift, deviation from target refraction). The influence of optic-haptic angulation proved to be significantly greater and more variable than edge design.

Acrylic Resins↗

Comparison of 4 methods for quantifying posterior capsule opacification.

PURPOSE: To compare the results of posterior capsule opacification (PCO) quantification and the repeatability of a fully automated analysis system (Automated Quantification of After-Cataract [AQUA]) with that of 2 other quantification methods and subjective grading of PCO. A test set of digital retroillumination images of 100 eyes with PCO of varying degrees was used. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: One hundred digital retroillumination images of eyes (100 patients) with PCO were selected to attain an even distribution from mild to severe cases. The images were evaluated by 4 methods: subjective grading by 4 experienced and 4 inexperienced examiners, the subjective Evaluation of Posterior Capsular Opacification (EPCO) system, posterior capsule opacification (POCO) software, and the AQUA system. Ten images were presented twice to assess the reproducibility of the analysis systems. RESULTS: Subjective grading correlated best with the subjective EPCO system and the objective AQUA system (r = 0.94 and r = 0.93, respectively). The POCO system showed very early saturation and therefore a much weaker correlation (r = 0.73). The POCO scores reached the maximum of 100% in several minimal to mild PCO cases. The reproducibility of the AQUA software was perfect and that of the other analysis systems, comparably satisfactory. CONCLUSION: The objective AQUA score correlated well with subjective methods including the EPCO system. The POCO system, which assesses PCO area, did not adequately describe PCO intensity and includes a subjective step in the analysis process. The AQUA system could become an important tool for randomized masked trials of PCO inhibition.

Cataract↗

Intraocular lens movement caused by ciliary muscle contraction.

PURPOSE: To investigate intraocular lens (IOL) movement, measured as a change in anterior chamber depth (ACD) caused by pilocarpine-induced ciliary muscle contraction. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: In this prospective study, the ACD was measured using high-precision, high-resolution, dual-beam partial coherence interferometry in 62 pseudophakic eyes of 55 patients under pilocarpine- and cyclopentolate-induced ciliary muscle contraction and relaxation. The following were studied: 2 models of a ring-haptic IOL (designed to accommodate), a plate-haptic IOL, and 3 types of 3-piece IOLs. Measurements were performed 3 months after surgery. RESULTS: The ring-haptic IOLs and plate-haptic IOL showed a forward movement (ring haptic 43A, -116 microm; ring haptic 43E, -222 microm; plate haptic -162 microm). The 3-piece IOLs showed no change in ACD except in 1 IOL type in which there was backward movement (156 microm). CONCLUSIONS: Pilocarpine-induced ciliary muscle contraction caused forward movement of ring- and plate-haptic IOLs that resulted in an estimated accommodative amplitude of less than 0.50 diopter in most cases. The accommodating ring-haptic IOLs did not perform better than the conventional plate-haptic IOL.

Accommodation, Ocular↗

Corneal endothelial cell protection with a dispersive viscoelastic material and an irrigating solution during phacoemulsification: low-cost versus expensive combination.

PURPOSE: To evaluate the protective effect on corneal endothelial cells of a low-cost and an expensive combination of a dispersive viscoelastic material and an irrigating solution during phacoemulsification. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This prospective randomized examiner- and patient-masked study comprised 90 eyes of 45 consecutive patients with age-related cataract in both eyes. For each patient, the first eye was randomly assigned to receive hydroxypropyl methylcellulose 2% (Ocucoat) and Ringer's solution (low-cost combination) or sodium chondroitin sulfate 4%-sodium hyaluronate 3% (Viscoat) and an enriched balanced salt solution (BSS Plus) (expensive combination) during phacoemulsification. The contralateral eye received the other treatment. Endothelial cell function was evaluated by measuring corneal thickness (CT) using partial coherence interferometry, morphology assessment, and endothelial cell counts. RESULTS: The acute postoperative increase in CT was +9.8 microm in the low-cost group and +10.9 microm in the expensive group; the difference between groups was not significant. After 1 month, the CT still differed significantly from baseline in the low-cost group. Three months after surgery, the CT had returned to baseline values in both groups. There was no significant between-group difference in endothelial cell counts or morphology. CONCLUSIONS: During phacoemulsification in a nonselected patient population, there was no difference in acute postoperative corneal edema and endothelial cell morphology after 3 months between a Viscoat and BSS Plus combination and an Ocucoat and Ringer's solution combination. Eyes receiving the expensive combination had marginally faster recovery of corneal swelling by 3 months. However, the cost of Viscoat and 500 mL BSS Plus is 5 times that of Ocucoat and Ringer's solution.

Aged↗

After-cataract in adults with primary posterior capsulorhexis: comparison of hydrogel and silicone intraocular lenses with round edges after 2 years.

PURPOSE: To evaluate the potential of primary posterior continuous curvilinear capsulorhexis (PCCC) to prevent after-cataract 2 years after surgery and perform a bilateral comparison of 2 intraocular lens (IOL) materials, silicone and hydrogel. SETTING: Department of Ophthalmology, University of Vienna, Medical School, Vienna, Austria. METHODS: In this randomized patient- and examiner-masked study, 37 patients (74 eyes) with age-related cataract had bilateral small-incision cataract surgery with PCCC. Randomly, 1 eye received a hydrogel IOL and the contralateral eye, a silicone IOL. Both IOLs had open-loop haptics and a round-edged optic. Standardized digital retroillumination photographs were taken 1 day and 1, 6, 12, and 24 months after surgery to evaluate development of after-cataract on the anterior and posterior capsules by subjective grading. Opacification of the PCCC area was also objectively evaluated using new software. RESULTS: Twenty-nine patients (58 eyes) completed the 2-year follow-up. Partial closure with ongrowth at the edge of the PCCC was found in 55% in the hydrogel group and 28% in the silicone group. Total closure of the PCCC was observed in 3 eyes, 2 in the hydrogel group and 1 in the silicone group. CONCLUSION: Ongrowth onto the PCCC area was dependent on IOL material. Because of less ongrowth and more fibrotic after-cataract with silicone IOLs, the efficacy of the PCCC was higher in this group.

Aged↗

Effect of a fixed dorzolamide-timolol combination on intraocular pressure after small-incision cataract surgery with Viscoat.

PURPOSE: To evaluate the effect of a fixed dorzolamide-timolol combination (Cosopt) on intraocular pressure (IOP) after small-incision cataract surgery with sodium chondroitin sulfate 4%-sodium hyaluronate 3% (Viscoat). SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This prospective randomized study comprised 76 eyes of 38 patients scheduled for small-incision cataract surgery in both eyes. Patients were randomized to receive 1 drop of the fixed dorzolamide-timolol combination in 1 eye or no treatment (control) immediately after cataract surgery. The fellow eye received the other assigned treatment. Cataract surgery was performed with Viscoat in an identical fashion in both eyes. The IOP was measured preoperatively and 6 hours, 20 to 24 hours, and 1 week postoperatively. RESULTS: Six hours after surgery, the mean increase in IOP was significantly lower in the dorzolamide-timolol group than in the control group (4.3 mm Hg +/- 5.6 [SD] versus 8.4 +/- 6.1 mm Hg; P =.003). Two eyes in the dorzolamide-timolol group and 9 in the control group had IOP spikes of 30 mm Hg or higher (P =.022). Twenty to 24 hours after surgery, the mean IOP change was -2.6 +/- 3.3 mm Hg in the dorzolamide-timolol group and 1.5 +/- 3.2 mm Hg in the control group (P<.001). CONCLUSIONS: The fixed dorzolamide-timolol combination was effective in reducing IOP 6 hours and 20 to 24 hours after cataract surgery. However, it did not prevent Viscoat-induced IOP spikes of 30 mm Hg or higher.

Aged↗

Postoperative change in effective lens position of a 3-piece acrylic intraocular lens.

PURPOSE: To study the change in postoperative position of an angulated, 3-piece, foldable acrylic intraocular lenses (IOLs) and evaluate the effect of a sharp posterior optic edge compared with that of a round optic edge on the change in postoperative anterior chamber depth (ACD). SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This randomized intraindividual-comparison study comprised 104 eyes of 52 patients with age-related cataract. Patient received a sharp-edged Sensar OptiEdge AR40e IOL (Allergan Surgical) in 1 eye and a round-edged Sensar AR40 IOL in the other eye. Postoperative follow-up included ACD measurement by partial coherence interferometry and evaluation of the capsulorhexis area by standardized retroillumination photography at 1 day, 1 week, and 1 and 6 months. RESULTS: A decrease in ACD during the first postoperative week was followed by a small increase in ACD during the first 6 months. There was no significant difference between the sharp-edged group and the round-edged group. CONCLUSIONS: The angulated 3-piece acrylic IOLs showed significant forward movement over the first postoperative 6 months. Although the change in refraction was small, there was variability among patients. The sharp posterior optic edge design did not affect the IOL's movement in the capsular bag.

Acrylic Resins↗

Influence of intraocular lens material on regeneratory posterior capsule opacification after neodymium:YAG laser capsulotomy.

PURPOSE: To evaluate the influence of a neodymium:YAG (Nd:YAG) laser capsulotomy on the morphology and development of regeneratory opacification of the remaining surrounding posterior capsule. SETTING: Department of Ophthalmology, Medical School, University of Vienna, Vienna, Austria. METHODS: Standardized digital retroillumination photographs were taken immediately before and after Nd:YAG laser capsulotomy and at 1 week, 6 months, and 1 to 3 years in 38 eyes of consecutive patients. Changes in regeneratory posterior capsule opacification (PCO) and opacification around the capsulotomy opening were evaluated. The influence of the type of intraocular lens (IOL) material and design (acrylic, n = 8; hydrogel, n = 8; silicone open loop, n = 11; silicone plate haptic, n = 6; poly(methyl methacrylate) [PMMA], n = 5) was assessed. The state of the anterior vitreous surface was examined at the slitlamp in all eyes. RESULTS: In 8 of 17 eyes with a silicone IOL, reduced regeneratory PCO was observed. Massive pearl formation on the margin of the Nd:YAG capsulotomy was typical with silicone IOLs (8 eyes) but also occurred with PMMA IOLs. Eyes with acrylic IOLs had no change in regeneratory PCO after the capsulotomy. Three of 8 eyes with hydrogel IOLs had complete closure of the posterior capsulotomy opening. CONCLUSIONS: Neodymium:YAG laser capsulotomy induced changes in the development and morphology of regeneratory PCO. Silicone and PMMA IOLs led to significant pearl formation on the capsulotomy margin, often combined with a reduction of peripheral regeneratory PCO (silicone). Hydrogel IOLs led to a higher incidence of reclosure of the Nd:YAG capsulotomy opening.

Acrylic Resins↗

Long-term effect of phacoemulsification on intraocular pressure after trabeculectomy.

PURPOSE: To evaluate the effect of temporal clear corneal phacoemulsification on intraocular pressure (IOP) in eyes that have had trabeculectomy. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This retrospective study evaluated the IOP in 48 eyes (35 patients) that had temporal clear corneal phacoemulsification after trabeculectomy. The mean interval between trabeculectomy and phacoemulsification was 27 months +/- 21 (SD) and the mean follow-up after phacoemulsification, 23 +/- 12 months. Intraocular pressure and antiglaucoma therapy before cataract surgery and at the end of follow-up were evaluated. For statistical analysis, the paired t test, Wilcoxon test, and chi-square test were used. RESULTS: Thirty-five eyes (73%) preoperatively and 25 eyes (52%) postoperatively were controlled (IOP < 22 mm Hg) without antiglaucoma therapy. The difference was statistically significant (P =.04, chi-square test). At the end of follow-up, the increase in mean IOP (1.6 mm Hg) and in mean number of antiglaucoma medications (0.4) was statistically significant (P =.002 and P =.05, respectively). CONCLUSIONS: Temporal clear corneal phacoemulsification after trabeculectomy was followed by a slight but statistically significant increase in IOP and the need for antiglaucoma medication after 2 years. However, the impairment in IOP control is comparable to that in the natural course of trabeculectomy.

Aged↗

Reproducibility of standardized retroillumination photography for quantification of posterior capsule opacification.

PURPOSE: To determine the short-term reproducibility of standardized digital retroillumination images of regeneratory posterior capsule opacification (PCO) using the photographic setup at 1 institution. SETTING: Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: In this prospective study, 60 retroillumination images of 30 eyes with varying degrees of PCO and different types of intraocular lenses were acquired with a standardized digital coaxial retroillumination system. Two images were taken per eye with a 1-minute interval between images. Ten other eyes were photographed in the same way but with a 5-day interval between the 2 images. All images were evaluated with a fully automated, objective PCO analysis software in which the PCO score was from 0 to 100. The 2 results (A, B) in each eye were compared, and the differences were calculated. RESULTS: There was a high correlation between the A and B results (r = 0.99). The mean absolute difference was 3.7%. The repeatability coefficient was 8.8%. CONCLUSION: Digital coaxial retroillumination photography provided quick acquisition of regeneratory PCO images. It provided excellent image quality and high reproducibility. The technique forms a good basis for automated quantification of PCO with new software systems.

Cataract↗

Biometry of cataractous eyes using partial coherence interferometry: clinical feasibility study of a commercial prototype I.

PURPOSE: To evaluate the clinical feasibility of the prototype version of a commercial partial coherence interferometry instrument (axial length measurement, ALM, Carl Zeiss Jena) for noninvasive, high-precision biometry in cataractous eyes. SETTING: Department of Ophthalmology, Vienna General Hospital, and Institute of Medical Physics, University of Vienna, Austria. METHODS: The preoperative axial length in 49 eyes of 37 cataract patients was measured with the commercial (ALM) and laboratory (PCI) prototypes of the partial coherence interferometry instrument, as well as with immersion ultrasound (IUS). RESULTS: Axial length measurements with the ALM and PCI did not differ significantly (P = .23). Both prototypes assessed longer axial lengths than the IUS technique (P < .0001; median 203 microm; range -476 to +635 microm). The precision of the axial length measurement was 18 microm, 28 microm, and 54 microm with the PCI, ALM, and IUS, respectively. CONCLUSIONS: Partial coherence tomography is a high-precision, high-resolution, noncontact biometric technique.The commercial PCI prototype is practical in clinical use, with improved comfort for patients, no need for anesthesia, and a reduced risk of infection. However, the difference between the PCI and IUS in axial length measurement must be considered when using the constants supplied by intraocular lens (IOL) manufacturers for IOL power calculations.

Adult↗

Assessment of anterior capsule opacification: photographic technique and quantification.

PURPOSE: To develop and evaluate a standardized slitlamp photographic technique to document and quantify anterior capsule opacification (ACO) in the pseudophakic eye. SETTING: University of Vienna, Medical School, Department of Ophthalmology, Vienna, Austria. METHODS: In this prospective study, slitlamp photographs were taken in 29 eyes of 19 patients with ACO of varying degrees and intraocular lenses of different materials. Digital slitlamp photographs were taken by 2 examiners using a standardized protocol. The intensity of anterior fibrosis was evaluated using subjective grading by 2 masked examiners and by image analysis. RESULTS: A standardized protocol was developed for documentation of ACO. The intraexaminer (or short-term) reproducibility (r = 0.96, P < .01) and interexaminer reproducibility (r = 0.93, P < .01) were excellent. The results of image analysis used to objectively quantify ACO correlated with the subjective grading (r = 0.95, P < .01). CONCLUSION: This standardized technique of slitlamp photography and image analysis provided reproducible documentation and can therefore serve as a basis for the quantification of ACO.

Anterior Chamber↗

Refractive outcome of cataract surgery using partial coherence interferometry and ultrasound biometry: clinical feasibility study of a commercial prototype II.

PURPOSE: To evaluate the refractive outcome of cataract patients 3 months postoperatively using optical biometry obtained with a prototype version (axial length measurement, ALM, Carl Zeiss Jena) of the commercial partial coherence interferometry (PCI) instrument (IOLMaster, Carl Zeiss Jena). SETTING: Department of Ophthalmology, Vienna General Hospital, and Institute of Medical Physics, University of Vienna, Austria. METHODS: Forty-five patients with age-related cataract in both eyes were scheduled for bilateral cataract surgery. Axial length was measured preoperatively with a prototype (ALM) of the commercial PCI instrument as well as with immersion ultrasound (IUS). Immersion US was performed by a single experienced investigator. In each patient, the first eye was randomly assigned to receive an intraocular lens (IOL) using the Holladay IOL power formula based on ALM or IUS biometry. The other biometric technique was used in the contralateral eye. Subjective refractive outcome was assessed 3 months postoperatively. RESULTS: Refractive outcomes with the 2 techniques did not differ significantly (P = .28). The mean numerical error (MNE) (the difference between the refractive outcome 3 months postoperatively and the predicted spherical equivalent) was 0.13 diopter (D) and 0.03 D for the ALM and IUS, respectively. The mean absolute error (MAE) (the absolute value of MNE) was 0.48 D (range 0.00 to 1.58 D) and 0.46 D (range 0.01 to 1.92 D) with the ALM and IUS, respectively. By recalculating the surgeon factor retrospectively to correct the Holladay formula to obtain a postoperative MNE of zero, a theoretical MAE of 0.46 D was obtained with both biometry techniques. CONCLUSIONS: Refractive outcome in cataract patients using PCI biometry was as good as that achieved with optimized IUS. However, the difference in axial length measured by the ALM compared to that measured by IUS must be considered when using the IOL A-constants supplied by the manufacturers.

Aged↗

Effect of an acrylic intraocular lens with a sharp posterior optic edge on posterior capsule opacification.

PURPOSE: To compare the posterior capsule opacification (PCO) inhibiting effect of the sharp posterior optic edge design of the Sensar OptiEdge AR40e intraocular lens (IOL) (Allergan Surgical) with that of the round-edged design of the Sensar AR40 IOL. Setting Department of Ophthalmology, University of Vienna, Vienna, Austria. METHODS: This prospective randomized patient- and examiner-masked study comprised 106 eyes of 53 patients with bilateral age-related cataract. Each patient had cataract surgery in both eyes and received an AR40 IOL in 1 eye and an AR40e IOL in the other eye. Postoperative examinations were at 1 week, 2 and 6 months, and 1 year. Digital slitlamp and digital retroillumination images of each eye were taken. The amount of PCO was assessed subjectively at the slitlamp and objectively using automated image-analysis software. RESULTS: The AR40e group had significantly less regeneratory and fibrotic PCO 1 year after surgery. The mean automated image-analysis software PCO score (scale 0 to 10) was 2.19 in the AR40 group and 1.10 in the AR40e group (P <.001). The AR40e group had less peripheral fibrotic PCO. There was no significant difference in patient-reported edge glare between the 2 IOL groups. CONCLUSIONS: The sharp-edged design of the Sensar OptiEdge AR40e IOL led to significantly less PCO than the round-edged AR40 IOL 1 year postoperatively.

Acrylic Resins↗