Short tandem repeat typing by capillary array electrophoresis: comparison of sizing accuracy and precision using different buffer systems.

Polymorphic microsatellite markers are widely used in gene discovery and mapping, human identification, agricultural genetics, and diagnosis of triplet-repeat expansion disorders. Reliable genotyping of these markers requires polymerase chain reaction (PCR) amplification and very-high-resolution electrophoresis. Capillary array electrophoresis offers extremely fast, high-resolution separation of DNA and more automated sample processing because labor-intensive slab-gel pouring and sample loading are eliminated. We report a simple, reliable procedure for preparing PCR samples for electrokinetic injection into capillaries using a 96-well tray and float dialysis. We developed an improved sizing standard for genotyping and used it to evaluate systematically the sizing accuracy and precision of low-viscosity, replaceable matrix formulations. Our study sizing over 28,000 alleles yielded an average precision of +/- 0.12 bp for fragments up to 350 bp. Low-viscosity formulations permit low-pressure matrix injection (40 psi) and a turnaround time of 70 min for 48-96 samples.

Alleles

Versatile low-viscosity sieving matrices for nondenaturing DNA separations using capillary array electrophoresis.

The high-resolution separation of double-stranded DNA (dsDNA) has important applications in physical mapping strategies and in the analysis of polymerase chain reaction (PCR) products. Although high-resolution separations of dsDNA by capillary electrophoresis (CE) have been reported, pulsed fields were required to achieve complete resolution of DNA fragments beyond 23 kilobase pairs (kbp). Here, we report a single formulation to separate a broad range (80 bp-40 kbp) of DNA fragments without the use of pulsed fields. We used a low-viscosity sieving medium (ca. 5 cP, at 25 degrees C) based on polyethyleneoxide (PEO) to separate DNA fragments up to 40 kbp. The matrix contained a mixture of 0.5% PEO (Mn 10(6)) to separate fragments up to 1.5 kbp, combined with 0.1% PEO (Mn 8 x 10(6)) to separate fragments between 1-40 kbp, within a single run. All PEO matrix formulations tested were compatible with a variety of intercalating dyes and with two different capillary wall coating methods. We obtained a detection limit of 25 fg of a 200 bp DNA quantitation standard using Vistra Green in the matrix. Resolution was best using short injection times (5 s or less) and low field strengths (approximately 100 V/cm). Sample runs were complete in 70 min, and use of the capillary array electrophoresis (CAE) system permitted high-throughput DNA analysis. The size range separated is approximately 10 times greater than with conventional slab gel separations.

Coloring Agents

Capillary zone electrophoresis of proteins.

This review article with 237 references is focused on capillary zone electrophoresis (CZE) of proteins. It includes discussion of modeling electrophoretic migration of proteins, sample pretreatment before the analysis, methods reducing the sorptions of proteins on the capillary wall, and techniques for increasing selectivity by using electrolyte additives including the sieving matrices. Significant progress in detection techniques, namely in laser-induced fluorescence and mass spectrometry, is emphasized. Modifications of CZE using specific interactions, such as affinity capillary electrophoresis or capillary immunoelectrophoresis, are debated as well as combination of CZE with other separation methods such as high performance liquid chromatography (HPLC). A number of practical applications of CZE of proteins are described.

Adsorption

Experimental evaluation of the separation efficiency in capillary electrophoresis using open tubular and gel-filled columns.

The band dispersion phenomena in capillary zone electrophoresis (CZE) using untreated and surface-treated open tubular and gel-filled capillaries were experimentally evaluated, with emphasis on small capillary diameters (10-100 microns). Laser-induced fluorescence detection was used for high-sensitivity detection of the isoindoles originated from model amino acids. The plots of plate height vs electric field strength were generated for different column radii and compared with a theoretical model for CZE. In addition to the diffusion-controlled band dispersion in the relatively low electric field range, adsorptive interactions between a solute and the capillary wall may play a certain role in band-broadening. The sorption-desorption kinetics become important with increasing electric field strength. Thermal effects appear to contribute little to band-broadening in relatively small capillaries (less than 50-microns i.d.) within normal operating voltages (less than 30 kV), but could become significant in capillaries with larger bores (greater than 75-microns i.d.). With gel-filled capillaries of small diameters (less than 50-microns i.d.), diffusion processes can be minimized. In addition, thermal effects do not appear critical in such columns at reasonable voltages.

Animals

Electrophoretic separations of proteins in capillaries with hydrolytically stable surface structures.

A procedure for obtaining highly stable coated capillaries for use in capillary electrophoresis (CE) is described. Reaction of surface-chlorinated fused silica capillaries with the Grignard reagent, vinyl magnesium bromide, followed by reaction of the vinyl group with acrylamide, results in an immobilized layer of polyacrylamide attached through hydrolytically stable Si-C bonds. This method is an extension of the capillary coating procedure described previously by Hjerten, differing in the means by which the polyacrylamide layer is bonded to the capillary walls. Capillaries treated in the manner described here can be used over a pH range of 2-10.5, without noticeable decomposition of the coating. In comparison to uncoated capillaries, separations of proteins using such coated capillaries are improved due to a reduction in protein adsorption to the capillary walls, although interaction is still present to some degree as evidenced by an inability to obtain plate counts as high as those predicted by theory. Electroosmotic flow is virtually eliminated in the coated capillaries, resulting in improved reproducibilities of protein migration times in comparison to uncoated capillaries. Additionally, peak skew is evaluated for model proteins and improvements are noted for the coated capillaries. Results are presented for separations of model protein mixtures, comparing the performance of the vinyl-bound polyacrylamide coated capillaries and uncoated capillaries at both high and low pH extremes.

Chemical Phenomena

Capillary zone electrophoresis of oligonucleotides. Factors affecting separation.

The influence of various parameters affecting separation of oligonucleotides by capillary zone electrophoresis has been examined. The effects of pH, ionic strength, and various additives, including highly charged cations such as spermine, were studied. Using polycytidines as model compounds, it was demonstrated that pH in the range of 5-8 and ionic strength in the range of 20-200 mmol/l do not influence the separation of oligonucleotides substantially. However, with the addition of spermine to the background electrolyte, migration order was inverted as the effective mobilities of the larger oligonucleotides were greatly decreased. With the addition of spermine and sodium dodecyl sulfate, the separation+ of these model oligonucleotides was also significantly affected. The best separation of a homologous series of polycytidines was obtained with a background electrolyte containing 60 mmol/l histidine, 30 mmol/l glutamic acid, 50 mmol/l sodium dodecyl sulfate and 3 mmol/l spermine.

Electrophoresis

High-throughput DNA sequencing on a capillary array electrophoresis system.

A capillary array electrophoresis apparatus capable of running and analyzing 48 DNA sequencing samples simultaneously has been constructed. The instrument uses a replaceable sieving buffer and incorporates a convenient method for introducing the buffer into the capillaries. Data from laser-induced fluorescence are collected as four separate images, one for each optical channel. The integrated data analysis software employs an open architecture that allows use of any DNA base-calling algorithm. DNA sequencing runs are completed in approx. 1 hr (approximately 500 bases), and instrument turnaround time between runs is less than 15 min. Overall, the instrument throughput is on the order of 720 templates/day, or 360,000 bases/day.

Animals