PubMed Health⌕ Search

Biomedical subjects

W H Al-Najjar

Publications and source records attributed to W H Al-Najjar.

2 recordsLinked to original sources

Tin shielding thicknesses for electrons.

The transmission of 6, 9, 12, 16, and 20 MeV electrons from a linear accelerator through tin and lead is studied at 0.5 cm depth in polystyrene. These measurements are performed using a 10 x 10 cm2 cone and extend well into bremsstrahlung region. The results show certain advantages of using tin to shield sensitive organs from electron beams. Tin is non-toxic and creates less additional bremsstrahlung even though it must be thicker than the equivalent lead shield.

Electrons↗

Dosimetric aspects of small circular fields of 10 MV photon beam.

Dosimetric data obtained from different measurement techniques of 10 MV small circular photon beams of 12.5, 20, 30, and 40 mm diameter were studied. The effects of lack of lateral electronic equilibrium and steep dose gradient were noticed even when measurements were performed with a small volume (0.02 cc) ionization chamber. From an examination of the results, it is concluded that an average value of data obtained from several carefully performed film measurements may be used for routine radiosurgery.

Models, Structural↗