Parameters for predicting electromagnetic lithotripter failure: quality assurance implications.

Abstract

Despite the extensive use of lithotripsy for treating renal and biliary calculi, there has been little data reported regarding the causes and manifestations of lithotripter failure. The clinical and service records for 145 consecutive treatments performed with the Siemens Lithostar Plus were reviewed. Service record analysis revealed eight failures of shock wave generation during a 10-month period. Six of these failures were subtle and still allowed shock wave generation. There were five in-line ultrasound probe failures during this period. The most useful clinical parameter for predicting lithotripter failure was reduced severity of sonographically evident cavitation bubbles during treatment. Lack of stone fragmentation and unexpectedly low analgesia requirements at high-power levels were less useful in predicting lithotripter failure. All clinical parameters suffered from nonspecificity. Preliminary experience, with an ongoing quality assurance program using a test object hydrophone, suggests this is a useful method of predicting lithotripter function and avoiding compromised treatments.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

W J Davros, B S Garra, J A Goldberg, L L Murphy, R K Zeman. 1992. Parameters for predicting electromagnetic lithotripter failure: quality assurance implications.. https://pubmed.ncbi.nlm.nih.gov/10147669/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Tip cooling effect and failure mechanism of field-emitting carbon nanotubes.

The cooling effect accompanying field electron emission has been considered for a single carbon nanotube (CNT) used as a field emission (FE) electron source. An improved model for the failure mechanism of field emitting CNTs has been proposed and validated. Our model predicts a maximum temperature (T-max) located at an interior point rather than the tip of the CNTs, and the failure of the CNT emitters tends to take place at the T-max point, inducing a segment by segment breakdown process. A combination of Joule heating and electrostatic force effect is proposed responsible for initiating the failure of the field emitting CNT and validated by in situ FE observation.

Equipment Failure↗

[Importance of an initial check-up for diagnostic radiology apparatus management].

The management of apparatus for diagnostic imaging is an important job for radiological technologists. In the last 10 years we have encountered 2227 malfunctions in 3652 days. We detected 163 cases at the initial check-up, which accounted for 7.3% of total cases. By performing an initial check-up, we detected one malfunction every 21 days and were able to keep the effects of mishaps to a minimum, prevent accidents, and obtain stable-quality images.

Equipment Failure↗