PubMed · 10602637
Pulse nebulization in pneumatic devices.
Abstract
Aerosols of a physiological salt solution and aqueous solutions of salbutamol, sodium cromoglicate, and dornase alfa were generated in a pneumatic nebulizer and analyzed in a system with controlled humidity of air as a carrier gas. Mass distribution of aerosol particles and yield of generation for pulse nebulization were measured. Pulsation of generation was realized with an attachment maintained by a computer program. Opening times of the valve were in the range 50-800 ms. The results indicate the possibility of improving aerosol particle delivery to the lung using a pulse generation system.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
L Gradoń, T R Sosnowski, Z Podolec. 1999. Pulse nebulization in pneumatic devices.. https://doi.org/10.1080/10803548.1999.11076409