Abstract
Phase contrast MR angiography (PC-MRA) depends on phase shifts caused by blood flow. Generally, PC sequences employ one VENC (velocity-encoding) value for each encoded spatial direction to optimize the signal in major vessels during peak systolic flow. We compared a mono-(30) with a multi-(20/30/45) VENC-PC-MRA technique in 10 patients with peripheral arterial occlusive disease. In all patients, the multi-VENC-PC sequence enhanced the vascular signal in vessels with very different flow velocities in one measurement. Large fields-of-view can be measured in a relatively short examination time to obtain an overview of the peripheral arterial system of the patient when contrast-enhanced MRA is not possible.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Goyen, L J Heuser. 2000. Improved peripheral MRA using multi-velocity-encoding phase contrast-enhanced MRA techniques.. https://doi.org/10.1080/028418500127345145