Calculating bone-lead measurement variance.
Abstract
The technique of (109)Cd-based X-ray fluorescence (XRF) measurements of lead in bone is well established. A paper by some XRF researchers [Gordon CL, et al. The Reproducibility of (109)Cd-based X-ray Fluorescence Measurements of Bone Lead. Environ Health Perspect 102:690-694 (1994)] presented the currently practiced method for calculating the variance of an in vivo measurement once a calibration line has been established. This paper corrects typographical errors in the method published by those authors; presents a crude estimate of the measurement error that can be acquired without computational peak fitting programs; and draws attention to the measurement error attributable to covariance, an important feature in the construct of the currently accepted method that is flawed under certain circumstances.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
A C Todd. 2000. Calculating bone-lead measurement variance.. https://doi.org/10.1289/ehp.00108383