PubMed Health⌕ Search

PubMed · 10969955

A new device for endoscopic third ventriculostomy.

Abstract

Since its description by Dandy in 1922, several techniques have been used to perform third ventriculostomy under endoscopic control. Except for the blunt technique, in which the endoscope is used by itself to create the opening in the floor of the third ventricle, the other techniques require more than one instrument to perforate the floor of the ventricle and enlarge the ventriculostomy. The new device described is a sterilizable modified forceps that allows both the opening of the floor and the enlargement of the ventriculostomy in a simple and effective way. The new device has the following characteristics: 1) the tip of the forceps is thin enough to allow the easy perforation of the floor of the ventricle; 2) the inner surface of the jaws is smooth to avoid catching vessels of the basal cistern; and 3) the outer surface of the jaws has indentations that catch the edges of the opening to prevent them from slipping along the instrument's jaws. The ventricle floor is opened by gentle pressure of the forceps, which is slowly opened so that the edges of the aperture are caught by the distal outer indentation of the jaws, leading to an approximately 4-mm opening of the floor. This device has been used successfully in 10 consecutive patients. This new device allows surgeons to perform third ventriculostomy under endoscopic control in a very simple, quick, and effective way, avoiding the need for additional single-use instruments.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Decq, C Le Guerinel, S Palfi, M Djindjian, Y Kéravel, J P Nguyen. 2000. A new device for endoscopic third ventriculostomy.. https://doi.org/10.3171/jns.2000.93.3.0509

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Endoscopic controlled intramedullary fracture reduction. A case report about closed fracture reduction under visual control.

Intramedullary bone endoscopy (IBE) was introduced as a new endoscopic technique in the year 2002 to achieve and manipulate the medullary canal of long bone under direct visual control. We report on a patient with a proximal femur fracture, where the new technique of IBE was used for intramedullary fracture reduction under visual control for the first time. Thus, the intraoperative use of an image intensifier was reduced.

Endoscopes↗

Removal of a pellet from the left orbital cavity by image-guided endoscopic navigation.

In the cranio-maxillofacial field, computer-aided surgery based on computed tomography data is becoming increasingly important. Complex surgical procedures can be performed utilizing preoperative imaging to provide real-time localization of surgical instruments in the surgical field. Image-data-based navigation plays an ever-increasing role in cases of minimal invasive surgery. The case is presented here of a 58-year-old male patient referred after a hunting accident with a pellet in the left orbit. After transferring the preoperatively acquired computed tomography data to a commercially available navigation system, the bullet was removed through the original wound canal using an intraoperatively calibrated high-resolution endoscope.

Endoscopes↗

[Manipulator assisted endoscope guidance in functional endoscopic sinus surgery: proof of concept].

BACKGROUND: Functional endoscopic sinus surgery (FESS) is characterized by single-handed preparation and guidance of the endoscope by the nondominant hand. This results in an additional extension of operation time by up to 15% and ergonomic deficits. The aim of this study is the conception of an automated assistance system for FESS in view of the following questions: (1) Which degree of surgical automation is suitable for FESS? (2) Which design is suitable? (3) What are the properties of the technical system (planning, time, accuracy, precision) of the selected system? (4) Does the system offer potential for a clinical application? METHODS: In all 49 FESS were analyzed for surgical workflows. Measurement of the maximum forces within FESS was performed with 40 trials on an anatomical model. Three different mechanical systems were used in ten FESS and evaluated using the ICCAS Human-Machine Evaluation Scale. For realization of automated endoscope guidance an engine-driven and -braked manipulator (PA10-6c, Mitsubishi, Japan) was used. The technical parameters determined were expenditure of time for the preoperative planning of workspace, surgical accuracy and precision of the intraoperative endoscope positioning, maximal forces, and time. RESULTS: Concept-conditioned instrument changes amount to an average of 41.1 and 18.9% (5.21 min) time requirement for each FESS side. Maximum forces on the mucous membrane during a conventional FESS were measured at 9.8 N (5.9-9.8). Usability of the mechanical endoscope holder was estimated in 18 of 20 cases to be inferior to the standard procedure. The time needed for segmenting the intranasal workspace was 15.2 min (10.0-23.0). The maximum deviation of the automatically driven endoscope from a planned position amounted to 0.85 mm (manually 4.64 mm). The maximum force was measured with 1.1 N in the z direction (manually 9.8 N). Automated guidance of the endoscope to an intranasal position needed 7.25 s (6.4-7.9); manually 12.64 s (5.9-43.0). CONCLUSION: Guidance of the endoscope for FESS by an automated motor-driven system is possible. The conception which is based on workflow analysis favors a system with automatic definition of the workspace and a manual movement of the endoscope. The examined system offers a potential for clinical application. Definition of the automation level and development of a man-machine interface is more important than selection or reconstruction of a special manipulator for endoscope guidance in FESS from a surgical point of view.

Endoscopes↗