Modified liquid junction interface for nonaqueous capillary electrophoresis-mass spectrometry.
Abstract
Electrospray ionization (ESI) is the most widely used ionization method in on-line coupling of capillary electrophoresis-mass spectrometry (CE-MS). The conventional coaxial sheath flow electrospray interface is currently being replaced by the more sensitive nanoelectrospray technique. The usual limitation of nanoelectrospray CE-MS interface has been its short lifetime caused by deterioration of the metal coating on the CE capillary terminus. This article describes an easy way to construct a more durable and sensitive nanospray interface for nonaqueous CE-MS. In this approach a very thin glass spray capillary (ca. 30 microm outer diameter) is partly inserted inside the CE capillary, the junction being surrounded by the electrolyte medium, which is in contact with the platinum electrode. The interface was tested with five pharmaceuticals: methadone, pentazocine, levorphanol, dihydrocodeine, and morphine. Detection limits ranged from 12 to 540 fmol. Separation efficiency and reproducibility were also studied. The CE current was found to be stable and the migration times were highly reproducible. All the CE separations were carried out in a nonaqueous background electrolyte solution.