Human cardiac imaging at 3 T using phased array coils.
Abstract
Using a two-element phased array receiver coil, single breath-hold, ECG gated cardiac images of signal-to-noise ratios up to 130 and contrast-to-noise ratios exceeding 35 between myocardium and blood were recorded at 3 T. At several locations within the myocardium, T*(2) and B(0) inhomogeneity were determined. Because of shorter T*(2) times and larger B(0) inhomogeneities attributable to enhanced susceptibility effects, real-time cardiac imaging, the use of spiral scans, and echo planar imaging are expected to be considerably more difficult at 3 T.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R Noeske, F Seifert, K H Rhein, H Rinneberg. 2000. Human cardiac imaging at 3 T using phased array coils.. https://doi.org/10.1002/1522-2594(200012)44%3A6%3C978%3A%3Aaid-mrm22%3E3.0.co%3B2-9