Fibre optical scanning with high resolution in thin-layer chromatography.

Abstract

In this paper a high-performance thin-layer chromatography (HPTLC) scanner is presented in which a special fibre arrangement is used as HPTLC plate scanning interface. Measurements are taken with a set of 50 fibres at a distance of 400 to 500 microm above the HPTLC plate. Spatial resolutions on the HPTLC plate of better than 160 microm are possible. It takes less than 2 min to scan 450 spectra simultaneously in a range of 198 to 610 nm. The basic improvement of the item is the use of highly transparent glass fibres which provide excellent transmission at 200 nm and the use of a special fibre arrangement for plate illumination and detection.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

B Spangenberg, K F Klein. 2000-11-17. Fibre optical scanning with high resolution in thin-layer chromatography.. https://doi.org/10.1016/s0021-9673(00)00880-3

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Monolithic porous polymer layer for the separation of peptides and proteins using thin-layer chromatography coupled with MALDI-TOF-MS.

Plates for thin-layer chromatography (TLC) with an attached layer of porous polymer monolith have been prepared and used for the separation of small molecules, peptides, and proteins. The 50-200-mum. thin poly(butyl methacrylate-co-ethylene dimethacrylate) layers were prepared in situ using UV-initiated polymerization. Precise control of the reaction conditions enables the preparation of monolithic layers with a well-defined porous structure that determines the chromatographic performance. Compared to conventional TLC and high-performance TLC using precoated layers based on silica, the small layer thickness and absence of any binder is expected to improve both retention characteristics and separation efficiency of the polymer-based monolithic thin-layer chromatographic plates. Spots of the separated compounds were first detected using typical UV imaging. Since the monolithic thin layers can be also prepared directly on the stainless steel MALDI carrier plate, the separation in TLC format can be coupled with MALDI-TOF-MS. Application of a conventional MALDI matrix facilitated desorption and ionization of peptides and proteins for molecular weight determination of the separated compounds.

Chromatography, Thin Layer↗

Separation of solanesol in tobacco leaves extract by slow rotary counter-current chromatography using a novel non-aqueous two-phase solvent system.

Non-aqueous solvent system composed of sunflower oil-ethanol was found to be suitable for the separation of lower polarity components. Employing this solvent system, 3 g of tobacco leaves extract containing 15% solanesol was separated using a slow rotary counter-current chromatograph equipped with an 1100 mL column made of 5.7 mm I.D. convoluted PTFE tube. The separation yielded 1.5 g of yellow oily product in which solanesol occupied 26.8% determined by HPLC. The tobacco tar and other main impurities in the tobacco leaves extract were removed after slow rotary counter-current chromatographic separation. The separation processing may be developed as a technique for producing a solanesol product with food grade.

Chromatography, Thin Layer↗

Selective thin-layer chromatography of 4-R-1,2,4-triazoles.

A convenient and selective TLC of 4-R-1,2,4-triazoles (Rtrz) and nitrogen containing aromatic rings is presented (molecules 1-11). This simple revelation method implies a complexation of heterocycles to Co(II) ions on a TLC plate, followed by a subsequent modification of the reduction potential which enables oxidation of Co(II) by permanganate.

Chromatography, Thin Layer↗