PubMed Health⌕ Search

PubMed · 11128163

Validation of a high-performance thin-layer chromatographic method for trace analysis for some generic drugs affecting gastrointestinal function.

Abstract

To prevent cross-contamination between pharmaceutical products manufactured with the same equipment, cleanup procedures must be introduced before the manufacture of a new product begins. From an analytical point of view, it is crucial to select and validate a suitable analytical method to determine contaminants in the rinse water, swabs, and the placebo of the next product. High performance thin-layer chromatography (HPTLC) was chosen in our laboratory for this purpose and was optimized to meet the requirements of trace determination. The method was validated in terms of the limit of detection, limit of quantitation (LOQ), linearity close to the LOQ, sample preparation from the swab media and from the placebo of the next product made with the same equipment (recovery), precision, selectivity (interference from the swab and placebo matrixes), resolution (from related compounds), and robustness. The HPTLC method was applied to 2 different generic drugs affecting gastrointestinal function--the water-soluble H2-receptor antagonist ranitidine hydrochloride (RHCl) and the water-insoluble choleretic drug ursodeoxycholic acid (UDCA). Chromatography was performed on silica plates by using toluene-methanol-diethylamine (9 + 1 + 1, v/v/v) and n-heptane-ethyl acetate-glacial acetic acid (5 + 5 + 1, v/v/v) as the mobile phases for RHCl and UDCA, respectively. RHCl was measured in situ at 320 nm, whereas the detection of UDCA was performed at 502 nm after postchromatographic derivatization. The method was used for the determination of RHCl and UDCA in the swabs, the final rinse water, and the placebo batch after the cleanup process.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J Novaković. Validation of a high-performance thin-layer chromatographic method for trace analysis for some generic drugs affecting gastrointestinal function.. https://pubmed.ncbi.nlm.nih.gov/11128163/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Monolithic porous polymer layer for the separation of peptides and proteins using thin-layer chromatography coupled with MALDI-TOF-MS.

Plates for thin-layer chromatography (TLC) with an attached layer of porous polymer monolith have been prepared and used for the separation of small molecules, peptides, and proteins. The 50-200-mum. thin poly(butyl methacrylate-co-ethylene dimethacrylate) layers were prepared in situ using UV-initiated polymerization. Precise control of the reaction conditions enables the preparation of monolithic layers with a well-defined porous structure that determines the chromatographic performance. Compared to conventional TLC and high-performance TLC using precoated layers based on silica, the small layer thickness and absence of any binder is expected to improve both retention characteristics and separation efficiency of the polymer-based monolithic thin-layer chromatographic plates. Spots of the separated compounds were first detected using typical UV imaging. Since the monolithic thin layers can be also prepared directly on the stainless steel MALDI carrier plate, the separation in TLC format can be coupled with MALDI-TOF-MS. Application of a conventional MALDI matrix facilitated desorption and ionization of peptides and proteins for molecular weight determination of the separated compounds.

Chromatography, Thin Layer↗

Separation of solanesol in tobacco leaves extract by slow rotary counter-current chromatography using a novel non-aqueous two-phase solvent system.

Non-aqueous solvent system composed of sunflower oil-ethanol was found to be suitable for the separation of lower polarity components. Employing this solvent system, 3 g of tobacco leaves extract containing 15% solanesol was separated using a slow rotary counter-current chromatograph equipped with an 1100 mL column made of 5.7 mm I.D. convoluted PTFE tube. The separation yielded 1.5 g of yellow oily product in which solanesol occupied 26.8% determined by HPLC. The tobacco tar and other main impurities in the tobacco leaves extract were removed after slow rotary counter-current chromatographic separation. The separation processing may be developed as a technique for producing a solanesol product with food grade.

Chromatography, Thin Layer↗

Selective thin-layer chromatography of 4-R-1,2,4-triazoles.

A convenient and selective TLC of 4-R-1,2,4-triazoles (Rtrz) and nitrogen containing aromatic rings is presented (molecules 1-11). This simple revelation method implies a complexation of heterocycles to Co(II) ions on a TLC plate, followed by a subsequent modification of the reduction potential which enables oxidation of Co(II) by permanganate.

Chromatography, Thin Layer↗