Generalized optimal current patterns and electrical safety in EIT.

Abstract

There are a number of constraints which limit the current and voltages which can be applied on a multiple drive electrical imaging system. One obvious constraint is to limit the maximum ohmic power dissipated in the body. Current patterns optimizing distinguishability with respect to this constraint are singular functions of the difference of transconductance matrices with respect to the power norm (the optimal currents of Isaacson). If one constrains the total current (L1 norm) the optimal patterns are pair drives. On the other hand if one constrains the maximum current on each drive electrode (an L(infinity) norm), the optimal patterns have each drive channel set to the maximum source or sink current value. In this paper we consider appropriate safety constraints and discuss how to find the optimal current patterns with those constraints.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

W R Lionheart, J Kaipio, C N McLeod. 2001. Generalized optimal current patterns and electrical safety in EIT.. https://doi.org/10.1088/0967-3334%2F22%2F1%2F311

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Batch electrophoretic cells with Eyring fluids: analysis of the hydrodynamics.

Mixing and dispersion phenomena caused by the carrier fluid in an electrophoretic cell is the main subject of this study. In particular, the effects of Joule heating on temperature and velocity profiles for Eyring-model fluids (EMF) are studied. The heat transfer is sequentially coupled with momentum transfer to derive an analytical expression for both the temperature and the velocity profiles. These results are then used to show the hydrodynamic behavior of the fluid in a batch electrophoretic cell. Furthermore, the results obtained are useful to compare with the fluid behavior of other carriers of different rheology, such as Newtonian fluids, power-law fluids, and viscoelastic fluids that obey the CEF model. The results show that EMF are potentially good carriers for relatively high Joule heat generation and therefore good candidates to control mixing inside the electrophoretic cell.

Electric Impedance↗

Investigation into the origin of the noise of surface electrodes.

In the recording of biomedical signals, a significant noise component is introduced by the electrode. The magnitude of this noise is considerably higher than the equivalent thermal noise from the electrode impedance. As the noise in surface electrodes limits the resolution of biopotential recordings, it is important to understand its origin. It was found that the noise mainly originates in the electrolyte-skin interface and that it is highly dependent on the electrode gel used and the skin properties of the test subject. Depending on skin treatment, magnitudes between 1 and 20 microVrms were measured among subjects. When the metal-electrolyte interface was allowed time to stabilise, electrodes of different metals measured face to face all showed a negligibly small noise magnitude (< 1 microVrms). In pre-gelled electrodes, where the metal-electrolyte interface has stabilised, no difference in noise properties was found between Ag-AgCl electrodes and other metals when measured on the skin. In subjects at rest, the contribution of EMG signals to the total noise level was shown to be negligibly small compared with the noise contribution of the electrolyte-skin interface. The magnitude of the noise of electrodes appeared to be inversely proportional to the square root of the area of the electrode on the skin.

Electric Impedance↗