Continuous concentration of bacteria in a microfluidic flow cell using electrokinetic techniques.
Abstract
A novel method for the concentration of bacterial solutions is presented that implements electrokinetic techniques, zone electrophoresis (ZE) and isoelectric focusing (IEF), in a microfluidic device. The method requires low power (< 3e-5 W) and can be performed continuously on a flowing stream. The device consists of two palladium electrodes held in a flow cell constructed from layers of polymeric film held together by a pressure-sensitive adhesive. Both ZE and IEF are performed with carrier-free solutions in devices in which the electrodes are in intimate contact with the sample fluid. IEF experiments were performed using natural pH gradients; no carrier ampholyte solution was required. Experiments performed in buffer alone resulted in significant electroosmotic flow. Pretreatment of the sample chamber with bleach followed by a concentrated solution of cationic detergent effectively suppressed electroosmotic flow.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
C R Cabrera, P Yager. 2001. Continuous concentration of bacteria in a microfluidic flow cell using electrokinetic techniques.. https://doi.org/10.1002/1522-2683(200101)22%3A2%3C355%3A%3Aaid-elps355%3E3.0.co%3B2-c