Optimized 3-D CT scan protocol for longitudinal morphological estimation in craniofacial surgery.

Abstract

Frequent three-dimensional computed tomography scanning may cause deterioration of the lenses of the eye, which are susceptible to x-rays. The authors performed an experimental study using a phantom head to establish an optimized three-dimensional computed tomography scan protocol for longitudinal morphological estimation in craniofacial surgery. Volume computed tomography scans were performed using a Hi-Speed Advantage SG CT scanner (GE Medical Systems, Milwaukee, WI) in the axial plane with a combination of scan parameters of varied values. The radiation doses induced by each scanning were measured using thermoluminescent dosimeter chips attached to the position of the lenses in the phantom. Two-dimensional images in the coronal plane and three-dimensional images of the osseous surface were generated from each accumulated data set. For each scan parameter, the images generated from data accumulated using different values were compared. The study showed that lens radiation dose increased with tube potential and was almost directly proportional to tube current and 1/pitch. The slice thickness did not affect lens radiation doses significantly. Images with good contrast resolution and low artifact level sufficient for estimating morphological changes were obtained using a low tube potential of 100 kVp and a tube current of 100 mA. In regard to z-axis spatial resolution, a slice thickness of less than 3 mm was required for precisely pointing out bony edges in the two-dimensional reformation images. These results led us to conclude that volume computed tomography for longitudinal examination in craniofacial surgery should be kept to a minimum frequency and performed using a low-dose technique, small slice thickness, and large pitch.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Hirabayashi, N Umamoto, M Tachi, Y Sugawara, A Sakurai, K Harii. 2001. Optimized 3-D CT scan protocol for longitudinal morphological estimation in craniofacial surgery.. https://doi.org/10.1097/00001665-200103000-00009

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2 µm × 2 µm resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗