PubMed · 11401074

The infant lung function model: a mechanical analogue to test infant lung function equipment.

Abstract

To facilitate international multicentre studies and quality control of infant pulmonary function measurements, the European Respiratory Society-American Thoracic Society (ERS-ATS) working group for infant lung function testing aims to develop specifications for standardized infant lung function equipment and software. However, a standardized test device is also needed to test whether existing infant lung function equipment is able to meet these requirements. The authors have built a "mechanical model baby" consisting of a linear pump which can reproduce prerecorded tidal flow waveforms with a precision of 0.5% (full stroke), enabling the simulation of tidal and forced flow patterns. This linear pump can be connected to a series of copper lung volumes (range 50-300 mL) with known time constants, so that lung volumes can be reproduced with a precision of +/-1% at frequencies 10-120bpm. Five airflow resistors were built using sinter material. When assessed using flows 0-300 mL.s(-1) all resistors showed a quasilinear pressure/ flow relationship, with slopes 1.0-5.6 kPa.L(-1).s. These resistances could be reproduced with a precision of +/-2.5%. The infant lung model can also be used to assess frequency responses of infant lung function equipment, since the pump is capable of delivering low amplitude volumes up to 20 Hz in a pseudorandom noise manner. In summary, based on error estimations, this infant lung model is able to test whether or not infant lung function equipment meets the requirements suggested by the European Respiratory Society-American Thoracic Society standardization group, that is: flow measurements within +/-2.5%, volume and resistance measurements within +/-5%, frequency response: magnitude attenuation <+/-10% and phase shift <+/-3 degrees at 10 Hz.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

U Frey, B Reinmann, J Stocks. 2001. The infant lung function model: a mechanical analogue to test infant lung function equipment.. https://doi.org/10.1183/09031936.01.17407550

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Correction of compliance and resistance altered by endotracheal tube leaks and non-linear pressure/volume-relationships.

Measurements of lung compliance (C) and resistance (R) are influenced by endotracheal tube leaks (ETTL) as well as non-linear pressure/volume relationships (P/V relationship). To keep C and R reliable, we developed an algorithm to mathematically correct inspiratory and expiratory volume (V) and flow. In this study, a ventilated lung model for non-linear P/V relationship with adjustment of an increasing ETTL was studied. In addition, the recordings (airway pressure, flow, and volume) of 21 infants (median weight: 1,220 g, range: 640-2,160 g, with a median leak size of 32%, range: 24-56%) were investigated. C and R were calculated continuously from the recordings of flow, volume, and airway pressure over time according to the changing volume. A method especially developed for the analysis of non-linear pressure-volume-relationship (APVNL) was employed. C and R affected by leaks were corrected applying the newly developed mathematical algorithm and compared with measurements without leakage. C could be corrected up to a leak of 80% and R up to 55% leak at half tidal V for the model with non-linear P/V-R. C and R without leak and after leak correction did not differ significantly in all infants where the APVNL method was applied (P > 0.05).

Airway Resistance↗

Parenchymal tethering, airway wall stiffness, and the dynamics of bronchoconstriction.

We do not yet have a good quantitative understanding of how the force-velocity properties of airway smooth muscle interact with the opposing loads of parenchymal tethering and airway wall stiffness to produce the dynamics of bronchoconstriction. We therefore developed a two-dimensional computational model of a dynamically narrowing airway embedded in uniformly elastic lung parenchyma and compared the predictions of the model to published measurements of airway resistance made in rats and rabbits during the development of bronchoconstriction following a bolus injection of methacholine. The model accurately reproduced the experimental time-courses of airway resistance as a function of both lung inflation pressure and tidal volume. The model also showed that the stiffness of the airway wall is similar in rats and rabbits, and significantly greater than that of the lung parenchyma. Our results indicate that the main features of the dynamical nature of bronchoconstriction in vivo can be understood in terms of the classic Hill force-velocity relationship operating against elastic loads provided by the surrounding lung parenchyma and an airway wall that is stiffer than the parenchyma.

Airway Resistance↗

Use of dynamic compliance for open lung positive end-expiratory pressure titration in an experimental study.

OBJECTIVE: We tested whether the continuous monitoring of dynamic compliance could become a useful bedside tool for detecting the beginning of collapse of a fully recruited lung. DESIGN: Prospective laboratory animal investigation. SETTING: Clinical physiology research laboratory, University of Uppsala, Sweden. SUBJECTS: Eight pigs submitted to repeated lung lavages. INTERVENTIONS: Lung recruitment maneuver, the effect of which was confirmed by predefined oxygenation, lung mechanics, and computed tomography scan criteria, was followed by a positive end-expiratory pressure (PEEP) reduction trial in a volume control mode with a tidal volume of 6 mL/kg. Every 10 mins, PEEP was reduced in steps of 2 cm H2O starting from 24 cm H2O. During PEEP reduction, lung collapse was defined by the maximum dynamic compliance value after which a first measurable decrease occurred. Open lung PEEP according to dynamic compliance was then defined as the level of PEEP before the point of collapse. This value was compared with oxygenation (Pao2) and CT scans. MEASUREMENTS AND MAIN RESULTS: Pao2 and dynamic compliance were monitored continuously, whereas computed tomography scans were obtained at the end of each pressure step. Collapse defined by dynamic compliance occurred at a PEEP of 14 cm H2O. This level coincided with the oxygenation-based collapse point when also shunt started to increase and occurred one step before the percentage of nonaerated tissue on the computed tomography exceeded 5%. Open lung PEEP was thus at 16 cm H2O, the level at which oxygenation and computed tomography scan confirmed a fully open, not yet collapsed lung condition. CONCLUSIONS: In this experimental model, the continuous monitoring of dynamic compliance identified the beginning of collapse after lung recruitment. These findings were confirmed by oxygenation and computed tomography scans. This method might become a valuable bedside tool for identifying the level of PEEP that prevents end-expiratory collapse.

Airway Resistance↗