PubMed · 11420990
A knowledge-based boundary delineation system for contrast ventriculograms.
Abstract
Automated left-ventricle (LV) boundary delineation from contrast ventriculograms has been studied for decades. Unfortunately, no accurate methods have ever been reported. A new knowledge based multistage method to automatically delineate the LV boundary at end diastole (ED) and end systole (ES) is discussed in this paper. It has a mean absolute boundary error or about 2 mm and an associated ejection fraction error of about 6%. The method makes extensive use of knowledge about LV shape and movement. The processing includes a multiimage pixel region classification, shape regression, and rejection classification. The method was trained and cross-validated tested on a database of 375 studies whose ED and ES boundary had been manually traced as the ground truth. The cross-validated results presented in this paper show that the accuracy is close to and slightly above the interobserver variability.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
L Sui, R M Haralick, F H Sheehan. 2001. A knowledge-based boundary delineation system for contrast ventriculograms.. https://doi.org/10.1109/4233.924802. https://pubmed.health/record/pubmed%3A11420990/a-knowledge-based-boundary-delineation-system-for-contrast-ventriculograms
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.