PubMed · 11506002
Field-deployable sniffer for 2,4-dinitrotoluene detection.
Abstract
A field-deployable instrument has been developed to detect low-level 2,4-dinitrotoluene (2,4-DNT) vapors. The system is based on previously developed artificial nose technology and employs an array of sensory materials attached to the distal tips of an optical fiber bundle. Both semiselective and nonspecific, cross-reactive sensors were employed. Each sensor within the array responds differentially to vapor exposure so the array's fluorescence response patterns are unique for each analyte. The instrument is computationally "trained" to discriminate target response patterns from nontarget and background environments. This detection system has been applied to detect 2,4-DNT, an analyte commonly detected on the soil surface above buried 2,4,6-trinitrotoluene (TNT) land mines, in spiked soil and aqueous and ground samples. The system has been characterized and demonstrated the ability to detect 120 ppb 2,4-DNT vapor in blind (unknown) humidified samples during a supervised field test.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K J Albert, M L Myrick, S B Brown, D L James, F P Milanovich, D R Walt. 2001-08-01. Field-deployable sniffer for 2,4-dinitrotoluene detection.. https://doi.org/10.1021/es010829t
Cite the original work for its findings. Save a collection to share your selection of sources.