PubMed HealthSearch

PubMed · 117007

Porous titanium endosseous dental implants in Rhesus monkeys: microradiography and histological evaluation.

Abstract

Artificial tooth roots with porous surface coatings were fabricated by sintering spherical powder of titanium alloy to solid cylindrical cores. The tooth roots were implanted subgingivally in healed mandibular premolar extraction sites of fifteen Rhesus monkeys. Supracrestal abutments were screwed into pretapped holes in the superior aspect of the primary subgingival stage four to eight weeks after implantation of the root. Clinical evaluations were performed monthly. Ten animals were sacrificed for histological evaluation of the functioning free standing implants. Of twenty-nine implants placed, three were lost and four were rated failures on the basis of histological evaluation. Postmortem evaluations revealed bone growth into the porous surface coating of the primary stage of all the implants. The most characteristic features which could be used to describe differences in the implant histology were the buccal and lingual crestal bone heights measured in relation to the root porosity. Twelve of sixteen implants had crestal bone heights within one millimeter of the superior aspect of the root. Four other implants displayed excessive bone recession, revealing as much as one half of the root porosity supracrestally. The four implant failures could be related to unfavorable features of recipient bone sites. The results demonstrate that the bone growth into the porous surface coatings of artificial tooth roots is an efficacious method of dental implant fixation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F A Young, M Spector, C H Kresch. 1979. Porous titanium endosseous dental implants in Rhesus monkeys: microradiography and histological evaluation.. https://doi.org/10.1002/jbm.820130604

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Future developments and applications of biomaterials: an overview.

It is recommended that the emphasis of biomaterials research and development for the future should be to achieve improved reliability. Use of increasing numbers of implants per year coupled with decreasing long term (greater than 5 years) success rates are resulting in progressively larger numbers of reparative implant operations. This trend can be altered by emphasizing three areas of R&D: 1) Studies of composite biomaterial systems offering unique combinations of biological surface behavior and substrate mechanical performance; 2) Investigate mechanisms of interfacial reactions so that long term responses of the host-implant can be predicted; 3) Develop long term predictive relationships for biomaterials reliability based upon interfacial reactions, biomechanics, fracture mechanics, fatigue testing, and retrieval analysis. Brief examples of efforts to develop undrestanding in these three areas are described using bioglass coated metal and bioglass coated alumina implants.

Alloys