PubMed Health⌕ Search

PubMed · 11741058

A clinically practical method of manually assessing polyethylene liner thickness.

Abstract

BACKGROUND: Most orthopaedists do not have access to contemporary computer-assisted radiographic techniques and therefore must use manual radiographic methods to assess polyethylene wear. The accuracy of most manual methods, however, has not been verified on clinical radiographs. In this study, we used manufacturer-developed wear templates to measure polyethylene wear and compared that technique with two other commonly used manual radiographic methods. Our purpose was to compare the accuracy of these techniques and thus determine their usefulness in assessing polyethylene wear in individual patients. METHODS: We analyzed seventeen polyethylene liners that were retrieved during revision operations after a mean of 12.0 years in situ. With use of digital calipers, we directly measured the true minimum polyethylene thickness of the explanted liner. We then measured the polyethylene thickness on anteroposterior pelvic radiographs that had been made before the revision. Three different manual techniques were used to evaluate the radiographs: the Dorr method, the Livermore method, and the newly described wear-template method. The minimum polyethylene thickness that was calculated with use of each of these methods was compared with the thickness as determined by direct measurement of the explanted liner. RESULTS: The mean error for the Dorr method (1.54 +/- 1.21 mm) was significantly greater than that for both the Livermore method (0.07 +/- 0.62 mm) and the wear-template method (-0.04 +/- 0.28 mm) (p < 0.01). The mean error for the Dorr method was significantly greater than zero (p < 0.01), indicating that this method consistently overestimated the true measurement. In contrast, the mean errors for the Livermore and wear-template methods were not significantly different from zero (p = 0.64 and 0.58, respectively), indicating that these methods did not consistently underestimate or overestimate the true measurement. However, the large standard deviations for all three methods suggest that the ranges in error are wide. The limits of agreement for the Livermore technique ranged from a 1.17-mm underestimation to a 1.31-mm overestimation of the true thickness of the polyethylene; those of the wear-template technique were smaller, ranging from a 0.60-mm underestimation to a 0.52-mm overestimation. CONCLUSION: The practicing orthopaedist needs an accurate and efficient method for determining the polyethylene thickness of modular acetabular components. We found that the template method best met this requirement because it involved a simple procedure: placing the template over the radiograph and measuring the distance between the femoral head and the inner surface of the metal shell. The other methods were less accurate, required additional information from the manufacturer, were more time-consuming, and were not as easy to demonstrate to the patient.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D Pollock, C J Sychterz, C A Engh. 2001. A clinically practical method of manually assessing polyethylene liner thickness.. https://doi.org/10.2106/00004623-200112000-00006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Classification and localization of acetabular labral tears.

OBJECTIVE: The purpose of this study was to compare the findings on hip MR arthrography (MRA) with the published MRA and arthroscopic classifications of hip labral tears and to evaluate a clock-face method for localizing hip labral tears. DESIGN/PATIENTS: We retrospectively reviewed 65 hip MRA studies with correlative hip arthroscopies. Each labrum was evaluated on MRA using the classification system of Czerny and an MRA modification of the Lage arthroscopic classification. In addition, each tear was localized on MRA by using a clock-face description where 6 o'clock was the transverse ligament and 3 o'clock was anterior. These MRA findings were then correlated with the arthroscopic findings using the clock-face method of localization and the Lage arthroscopic classification of labral tears. RESULTS: At MRA, there were 42 Czerny grade 2 and 23 grade 3 labral tears and 22 MRA Lage type 1, 11 type 2, 22 type 3 and 10 type 4 tears. At arthroscopy, there were 10 Lage type 1 flap tears, 20 Lage type 2 fibrillated tears, 18 Lage type 3 longitudinal peripheral tears and 17 Lage type 4 unstable tears. The Czerny MRA classification and the modified MRA Lage classification had borderline correlation with the arthroscopic Lage classification. Localization of the tears using a clock-face description was within 1 o'clock of the arthroscopic localization of the tears in 85% of the patients. CONCLUSIONS: The Lage classification, which is the only published arthroscopic classification system for hip labral tears, does not correlate well with the Czerny MRA or an MRA modification of the Lage classification. Using a clock-face description to localize tears provides a way to accurately localize a labral tear and define its extent.

Acetabulum↗

Acetabular revisions using a cementless oblong cup: five to ten year results.

The purpose of this paper was to evaluate the results of acetabular revisions with the use of an oblong revision cup that is designed with its longitudinal diameter elongated relative to its transverse diameter. Between 1996 and 2001, 62 hips in 60 patients underwent an acetabular revision with the insertion of a LOR acetabular component. Seven hips were lost to follow-up or the patients died; the remaining 55 hips (53 patients) remained in follow-up for an average of 7.2 years (range: 5.0-10.1 years). One socket was revised for aseptic loosening, and another was operated on for a late polyethylene liner dissociation. The average Harris hip score (HHS) improved from 34 to 79. Results were rated as excellent in 16 hips, good in 28, fair in six and poor in three. Radiographic analysis demonstrated an improvement in the average vertical displacement of the hip centre: 49 hips had a well-fixed, bone-ingrown cup and four had a stable fibrous union. For large superolateral acetabular bone deficiencies, this implant facilitated a complex reconstruction without the need for bulk structural acetabular bone grafts, provided good clinical results and showed satisfactory stability at the midterm follow-up.

Acetabulum↗

Corona mortis: an anatomical study with clinical implications in approaches to the pelvis and acetabulum.

The "corona mortis" is an anatomical variant, an anastomosis between the obturator and the external iliac or inferior epigastric arteries or veins. It is located behind the superior pubic ramus at a variable distance from the symphysis pubis (range 40-96 mm). The name "corona mortis" or crown of death testifies to the importance of this feature, as significant hemorrhage may occur if accidentally cut and it is difficult to achieve subsequent hemostasis. It constitutes a hazard for orthopedic surgeons especially in the anterior approach to the acetabulum. We carried out forty cadaver dissections (80 hemi-pelvises) through the ilioinguinal approach. A vascular anastomosis was found in 83% of specimens. Of these, 60% had a large diameter (>3 mm) channel along the posterior aspect of the superior pubic ramus. In clinical practice, however, 492 anterior approaches (to the best of our knowledge the largest series described) have been carried out over the last 15 years by the senior author (MB) and only five of these problematic vessels were discovered, and in only two cases was there troublesome bleeding. This study confirms a paradox: in anatomical dissections a large vessel was identified behind the superior pubic ramus, whereas in clinical practice this vessel does not seem to be as great a threat as initially perceived. Orthopedic surgeons planning an anterior approach to the acetabulum, such as the ilioinguinal or the intrapelvic approach (modified Stoppa), have to be cautious when dissecting near the superior pubic ramus. Despite the high prevalence of these large retropubic vessels in the dissecting room, surgeons should exercise caution but not alter their surgical approach for fear of excessive hemorrhage.

Acetabulum↗