Universal standardization of forearm bone densitometry.

Abstract

As part of an effort to quantify device-dependent differences in forearm bone density, 101 women, aged 20-80 years (approximately 16 women in each age decade), were scanned on six forearm bone densitometers: the Aloka DCS-600EX, the Hologic QDR-4500A, the Lunar PIXI, the Norland pDEXA, the Osteometer DTX-200, and the Pronosco X-posure System. Regression statistics are reported for all similar regions of interest (ROIs). However, comparisons were confounded because of large differences in the ROI size and placement. The number of ROIs reported for a single scan by each device varied from 1 to 12. The correlation coefficients ranged from 0.7 < r < 0.97, with the highest correlation coefficients and lowest SEs for comparisons between the most similar ROIs. Standardized units of bone mineral density are derived for distal (sdBMD), mid-(smBMD), and proximal (spBMD) ROTs that allow for comparable mean bone densities to be derived for patient populations. Five phantoms were scanned and characterized on five of the devices and the precision and mean values were reported. These phantom values will aid in the in vitro cross-calibration between manufacturers to recreate the presented in vivo relationships. Care should be exercised when using these equations for cross-calibrating patient databases or pooling clinical data from different devices because the least significant differences detectable from measurements taken on two different machines can be increased substantially.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

John A Shepherd, Xiao Guang Cheng, Ying Lu, Chris Njeh, Jorg Toschke, Klaus Engelke, Michael Grigorian, Harry K Genant. 2002. Universal standardization of forearm bone densitometry.. https://doi.org/10.1359/jbmr.2002.17.4.734

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A repeated measures concordance correlation coefficient.

The concordance correlation coefficient is commonly used to assess agreement between two raters or two methods of measuring a response when the data are measured on a continuous scale. However, the situation may arise in which repeated measurements are taken for each rater or method, e.g. longitudinal studies in clinical trials or bioassay data with subsamples. This paper proposes a coefficient for measuring agreement between two raters or two methods of measuring a response in the presence of repeated measurements. We illustrate the methodology with examples comparing (1) 1-hr versus 2-hr blood draws for measuring cortisol in an asthma clinical trial and (2) two measurements of percentage body fat, from skinfold calipers and dual energy X-ray absorptiometry.

Absorptiometry, Photon↗

Crystallization and preliminary X-ray analysis of a domain in the Runx2 transcription factor that interacts with the 1alpha,25 dihydroxy vitamin D3 receptor.

The Runx2 transcription factor is a key regulator of osteoblast differentiation. In response to 1alpha,25 dihydroxy vitamin D3, Runx2 may interact with the 1alpha,25 dihydroxy vitamin D3 receptor (VDR) in the promoter of target genes, producing a synergic activation of their transcription. Previous studies have suggested that the motifs responsible for the VDR-Runx2 interaction are contained within the 230-361 domain of Runx2. In this work, we confirmed by GST-pull down that Runx2(I(209-361)) is sufficient to interact with the VDR. To obtain structural information, GST-Runx2(I(209-361)) protein was overexpressed in Escherichia coli, purified and crystallized using the hanging-drop vapor-diffusion method and polyethyleneglycol as a precipitant. The crystals were found to diffract to a maximum resolution of 2.7 A and a complete data set to a 3.3 A resolution was collected and analyzed. The crystals belong to the tetragonal system, with a space group P4 and unit-cell parameters of a = b = 90.8, and c = 57.2 A. The presence of a monomer of the recombinant GST-Runx2(I(209-361)) in the asymmetric unit gives a V(M) of 2.7 A(3) Da(-1) and a solvent content of 54.8%.

Absorptiometry, Photon↗

Quantifying the treatment effect explained by markers in the presence of measurement error.

Surrogate markers or intermediate markers are important in identifying subjects with high risk of a serious disease or for monitoring disease progression of a subject on treatment. Quantifying the proportion of treatment effect (PTE) explained by markers has been studied extensively. Due to reasons such as biological variation, limited machine precision, etc. markers are generally measured with error. The estimated PTE ignoring the measurement error could be biased, which may lead to incorrect conclusions. In this article, we adjust for the measurement error using regression calibration to construct a less biased estimator of excess relative odds, a quantity to measure the treatment effect explained by markers. The method is applied to data from a clinical study in osteoporosis.

Absorptiometry, Photon↗