Abstract
The present study is concerning the construction of ferricyanide-mediated Gluconobacter oxydans cell ethanol biosensor. The size exclusion effect of a cellulose acetate membrane was used for elimination of glucose interferences during ethanol assays in real samples. A typical response time of the biosensor was 13 s with a high sensitivity of 3.5 microA mM(-1). The microbial biosensor exhibits a very low detection limit of 0.85 microM and a wide linear range from 2 to 270 microM. The operational stability was excellent. During 8.5 h of repetitive ethanol assays, no decrease in the sensor sensitivity was observed. The biosensor was successfully used in the off-line monitoring of ethanol fermentation with a good agreement with HPLC measurements (R(2)=0.998).