PubMed · 12353289
Asymmetrical gradient coil for head imaging.
Abstract
This work presents a novel approach to develop dedicated transverse gradient coils for head imaging. The proposed coil design is based on the stochastic optimization of an asymmetrical stream function and improves the matching between the region-of-interest and the homogeneous gradient volume. Additionally, the electric field produced by these asymmetrical coils is 30% lower than that produced by standard symmetrical designs, which minimizes the risk of magnetostimulation of nerves in fast imaging techniques. A prototype of the asymmetrical gradient coil was built to test the method and magnetic field produced by the prototype was measured. Magnetic field measurements and electrical parameters of coils are in good agreement with theoretical calculations.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D Tomasi, R F Xavier, B Foerster, H Panepucci, A Tannús, E L Vidoto. 2002. Asymmetrical gradient coil for head imaging.. https://doi.org/10.1002/mrm.10263. https://pubmed.health/record/pubmed%3A12353289/asymmetrical-gradient-coil-for-head-imaging
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.