Multi-detector row CT: radiation dose characteristics.

Abstract

PURPOSE: To determine the dose characteristics of multi-detector row computed tomography (CT) and to provide tabulated dose values and rules of thumb that assist in minimizing the radiation dose at multi-detector row CT. MATERIALS AND METHODS: Weighted CT dose index (CTDI100w) values were obtained from three multi-detector row CT scanners (LightSpeed; GE Medical Systems, Milwaukee, Wis) for both head and body CT modes by using standard CT-dose phantoms. The CTDI100w was determined as a function of x-ray tube voltage (80, 100, 120, 140 kVp), tube current (range, 50-380 mA), tube rotation time (0.5-4.0 seconds), radiation profile width (RPW) (5, 10, 15, 20 mm), and acquisition mode (helical high-quality and high-speed modes and axial one-, two-, and four-section modes). Statistical regression was performed to characterize the relationships between CTDI100w and various technique factors. RESULTS: The CTDI100w (milligray) increased linearly with tube current: in head mode, CTDI100w = (0.391 mGy/mA +/- 0.004) x tube current (milliampere) (r2 = 0.999); in body mode, CTDI100w = (0.162 mGy/mA +/- 0.002) x tube current (milliampere) (r2 = 0.999). The CTDI100w increased linearly with rotation time: in head mode, CTDI100w = (34.7 mGy/sec +/- 0.2) x rotation time (seconds) (r2 = 1.0); in body mode, CTDI100w = (13.957 mGy/sec +/- 0.005) x rotation time (seconds) (r2 = 1.0). The relationship of normalized CTDI100w (milligrays per 100 mAs) with tube voltage followed a power law: in head mode, CTDI100w = (0.00016 mGy/100 mAs. kVp +/- 0.00007) x (tube voltage)(2.5+/-0.1) (r2 = 0.997); in body mode, CTDI100w = (0.000012 mGy/100 mAs. kVp +/- 0.000007) x (tube voltage)(2.8+/-0.1) (r2 = 0.996). In all scanning modes, CTDI100w decreased when RPW increased. CTDI100w was 10% higher in head mode and 13% lower in body mode compared with the value suggested by the manufacturer, which is displayed at the scanner console. When deposited power exceeded 24 kW, CTDI100w increased by 10% as a result of use of the large focal spot. CONCLUSION: The authors provide a set of tables of radiation dose as a function of imaging protocol to facilitate implementation of radiation dose-efficient studies.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Leena M Hamberg, James T Rhea, George J Hunter, James H Thrall. 2003. Multi-detector row CT: radiation dose characteristics.. https://doi.org/10.1148/radiol.2263020205

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Multidetector CT and dentascan software: dosimetric evaluation and technique improvement.

PURPOSE: The development of new operative techniques in oral and maxillofacial surgery within the last few years has led to an increasing demand for Dentascan examination, also in paediatric patients. It is necessary to modify acquisition parameters to reduce the absorbed dose. The aim of this study was to define a Dentascan protocol in which a reduced X-ray dose could be used. MATERIALS AND METHODS: Dosimeters were applied to the eyes, mouth, parotid glands, thyroid and back of the neck of an anthropomorphic Plexiglas phantom that underwent multidetector computed tomography (MDCT) Dentascan examinations. Both 120kV and 80 kV were used to study the mandibular and maxillary arches. RESULTS: Examinations obtained with the 80 kV protocol showed a ten-fold reduction in the absorbed dose, without affecting image quality. CONCLUSIONS: We suggest a Dentascan protocol that reduces the X-ray dose administered to the patient while ensuring the same high diagnostic accuracy.

Head↗