PubMed · 1263554
Interference with function of unipolar pacemaker due to muscle potentials.
Abstract
One hundred and seventy-three unipolar pulse generator systems were tested in 169 patients to determine the incidence of sensing of muscle potentials. One hundred and twenty-two implants in the pectoral region were suitable for assessment and 60 (49 per cent) demonstrated sensing. Ventricular-inhibited units sensed muscle potentials in 69 per cent (47 of 78), with 9 patients complaining of related symptoms. Ventricular-synchronous units sensed in 30 per cent (13 of 34) with no patient manifesting symptoms.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S T Anderson, B B Davis. 1976. Interference with function of unipolar pacemaker due to muscle potentials.. https://pubmed.ncbi.nlm.nih.gov/1263554/