Validation of dynamic MLC-controller log files using a two-dimensional diode array.

Abstract

Intensity-modulated radiation therapy (IMRT) delivered with multi-leaf collimator (MLC) in the step-and-shoot mode uses multiple static MLC segments to achieve intensity modulation. For typical IMRT treatment plans, significant numbers of segments are delivered with monitor units (MUs) of much less than 10. Verification of the ability of the linear accelerator (linac) to deliver small MU segments accurately is an important step in the IMRT commissioning and quality assurance (QA) process. Recent studies have reported large discrepancies between the intended and delivered segment MUs. These discrepancies could potentially cause large errors in the delivered patient dose. We have undertaken a systematic study to evaluate the accuracy of the dynamic MLC log files, which are created automatically by our commercial MLC workstation after each delivery, in recording the fractional MU delivered in the step-and-shoot mode. Two linac models were evaluated with simple-geometry leaf sequences and delivered with different total MUs and different nominal dose rates. A commercial two-dimensional diode array was used for the measurement. Large discrepancies between the intended and delivered segment MUs were found. The discrepancies were larger for small MU segments at higher dose rate, with some small MU segments completely undelivered. The recorded fractional MUs in the log files were found to agree with what was delivered within the limits of our experimental uncertainty. Our results indicate that it is important to verify the delivery accuracy of small MU segments that could potentially occur in a patient treatment and that the log files are useful in checking the integrity of the linac delivery once validated. Thus validated log files can be used as a QA tool for general IMRT delivery and patient-specific plan verification.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jonathan G Li, James F Dempsey, Li Ding, Chihray Liu, Jatinder R Palta. 2003. Validation of dynamic MLC-controller log files using a two-dimensional diode array.. https://doi.org/10.1118/1.1567951

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[Erroneous and unsatisfactory filling in of drug charts--a potential source of medication error].

BACKGROUND: Medication errors may have serious consequences for patients and erroneous drug charts represent one of several causes of error. As part of a quality assurance survey, we studied whether user documentation on the charts was in accordance with the guidelines for chart keeping. MATERIAL AND METHODS: On the same day, the drug charts of all hospitalized patients (n = 401) were copied and a representative sample of 207 charts was reviewed. RESULTS: 15 charts had no prescriptions; 14 of the remaining 192 charts (7 %) were correctly filled in. 152 charts (79 %) had deviations that we considered might pose only minor potential health risk; 26 (14 %) had deviations implying potential risk. No charts had deviations considered to pose major risk. INTERPRETATION: Several deviations from the guidelines were found and better training is needed; we recommend standardisation of drug charts and guidelines.

Documentation↗