Magnetic induction tomography: phase versus vector-voltmeter measurement techniques.
Abstract
In magnetic induction tomography (MIT) the in-quadrature component, and hence the phase, of the received signal contains information about the conductivity of the tissue. The quality of imaging will depend on the precision with which phase can be measured. Preliminary studies suggest that a precision of 10 m degrees may be required for a practical biomedical MIT system operating at 10 MHz. This paper describes the results of measurements carried out with a 16-channel, downconverting, 10 MHz, MIT system utilizing two types of data extraction techniques: direct-phase measurement and measurement of the in-phase and in-quadrature components of the signal with a vector voltmeter. The basic precision provided by each technique was 50 m degrees, with thermal drift representing the major limiting factor. Preliminary measurements of average conductivity and permittivity for a human thigh in vivo are given.