PubMed · 12964765
Measurement of enzyme kinetics using a continuous-flow microfluidic system.
Abstract
This paper describes a microanalytical method for determining enzyme kinetics using a continuous-flow microfluidic system. The analysis is carried out by immobilizing the enzyme on microbeads, packing the microbeads into a chip-based microreactor (volume approximately 1.0 nL), and flowing the substrate over the packed bed. Data were analyzed using the Lilly-Hornby equation and compared to values obtained from conventional measurements based on the Michaelis-Menten equation. The two different enzyme-catalyzed reactions studied were chosen so that the substrate would be nonfluorescent and the product fluorescent. The first reaction involved the horseradish peroxidase-catalyzed reaction between hydrogen peroxide and N-acetyl-3,7-dihydroxyphenoxazine (amplex red) to yield fluorescent resorufin, and the second the beta-galactosidase-catalyzed reaction of nonfluorescent resorufin-beta-D-galactopyranoside to yield D-galactose and fluorescent resorufin. In both cases, the microfluidics-based method yielded the same result obtained from the standard Michaelis-Menten treatment. The continuous-flow method required approximately 10 microL of substrate solution and 10(9) enzyme molecules. This approach provides a new means for rapid determination of enzyme kinetics in microfluidic systems, which may be useful for clinical diagnostics, and drug discovery and screening.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Gi Hun Seong, Jinseok Heo, Richard M Crooks. 2003-07-01. Measurement of enzyme kinetics using a continuous-flow microfluidic system.. https://doi.org/10.1021/ac034155b
Cite the original work for its findings. Save a collection to share your selection of sources.