The performance of the probability density function in differentiating supraventricular from ventricular rhythms.

Abstract

The ability of the probability density function (PDF) of an automatic implantable cardioverter defibrillator (AICD) to reject supraventricular arrhythmias being recognized as ventricular was evaluated in 12 patients who were treated with an AICD (Ventak P 1600). The PDF criterion was monitored via telemetry with the sinus rate during exercise test. PDF was satisfied in seven patients at a rate of 75-144/min (mean 109/min), and not in the remaining five patients (mean rate 141/min). PDF was fulfilled in five of ten patients at a lower heart rate than predicted by the duty cycle index, derived from the ventricular patch lead electrogram at the implantation. Thus PDF is often fulfilled already at a moderately elevated sinus rate. If used to prevent inadvertent AICD discharges during rapid supraventricular rhythms, its performance should be tested in the individual patient.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

L Toivonen, M Viitasalo, A Järvinen. 1992. The performance of the probability density function in differentiating supraventricular from ventricular rhythms.. https://doi.org/10.1111/j.1540-8159.1992.tb06837.x. https://pubmed.health/record/pubmed%3A1382273/the-performance-of-the-probability-density-function-in-differentiating-supraventricular-from-ventric

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related citations

A biventricular ICD system with biventricular defibrillation.

We describe the case of a 59-year-old gentleman with severe dilated cardiomyopathy requiring implantation of a dual-chamber biventricular implantable cardioverter-defibrillator (ICD). High defibrillation thresholds (DFT) were encountered at implant with an inadequate defibrillation safety margin. Testing of all possible shock vectors/polarities with and without the SVC coil and optimization of the distal RV coil position all proved inadequate. A satisfactory defibrillation safety margin was achieved following placement of a second lead in the coronary sinus to enable biventricular defibrillation. This case highlights an additional strategy for combating high DFTs and is an option even in dual-chamber biventricular ICD systems.

Electric Countershock↗