Two isocratic liquid chromatographic methods (conventional and micellar) for the determination of danazol (DZ) in capsules using canrenone (CAN) as internal standard have been developed and validated. In conventional liquid chromatography a mobile phase 35% water:acetonitrile 65%, v:v, a flow-rate 1 ml min(-1) and a C18 Hypersil ODS (250 x 4.6 mm, 5 microm) column (25 degrees C) were used. In micellar liquid chromatography (MLC) the conditions were: mobile phase 40 mM sodium dodecyl sulfate:2% pentanol, flow-rate 0.5 ml min(-1) and C18 Hypersil ODS (150 x 3.0 mm, 5 microm) column (60 degrees C). For both methods. UV absorbance detection at 280 nm was used and a separation up to base line was achieved. Prior to HPLC analysis a simple sample preparation was required. The recoveries found in the accuracy test were 99 +/- 10 and 101 +/- 8%, in conventional liquid chromatography (CLC) and MLC, respectively. Repeatability and intermediate precision expressed as R.S.D. were lower than 5% for both methods. Detection limits obtained were 2.4 and 3.0 ng g(-1) in CLC and CLM, respectively.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R Gonzalo-Lumbreras, R Izquierdo-Hornillos. 2003-07-14. Conventional and micellar liquid chromatography method development for danazol and validation in capsules.. https://doi.org/10.1016/s0731-7085(03)00224-3

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Development and validation of an HPLC method for the determination of spironolactone and its metabolites in paediatric plasma samples.

An HPLC method has been developed and validated for the determination of spironolactone, 7 alpha-thiomethylspirolactone and canrenone in paediatric plasma samples. The method utilises 200 microl of plasma and sample preparation involves protein precipitation followed by Solid Phase Extraction (SPE). Determination of standard curves of peak height ratio (PHR) against concentration was performed by weighted least squares linear regression using a weighting factor of 1/concentration2. The developed method was found to be linear over concentration ranges of 30-1000 ng/ml for spironolactone and 25-1000 ng/ml for 7 alpha-thiomethylspirolactone and canrenone. The lower limit of quantification for spironolactone, 7 alpha-thiomethylspirolactone and canrenone were calculated as 28, 20 and 25 ng/ml, respectively. The method was shown to be applicable to the determination of spironolactone, 7 alpha-thiomethylspirolactone and canrenone in paediatric plasma samples and also plasma from healthy human volunteers.

Canrenone↗

Use of partial least-squares regression for multicomponent determinations based on kinetic spectrofluorimetric data. Simultaneous determination of canrenone and spironolactone in urine.

A new spectrofluorimetric method for the simultaneous determination of canrenone and spironolactone in urine is proposed. The method is based on the different rates at which the two analytes react with hot sulfuric acid to form a trienone. The kinetic spectrofluorimetric data are processed by partial least-squares regression. The effects of sulfuric acid concentration and temperature on the system under study were also evaluated and the optimum values for carring out the reaction were 50% and 50 degrees C, respectively. The method was checked by analyzing urine samples that they contained both diuretics. The accuracy and the precision of the method were tested. The relative standard errors in the quantification of each analyte in all tested samples were 3.69 and 3.59%. The proposed method was validated by comparison with a high performance liquid chromatographic method for urine samples.

Canrenone↗