Numerical modeling of needle tip artifacts in MR gradient echo imaging.

Abstract

Exact determination of needle tip position is obsolete for interventional procedures under control of magnetic resonance imaging (MRI). Exact needle tip navigation is complicated by the paramagnetism of microsurgical instruments: Local magnetic field inhomogeneities are induced resulting in position encoding artifacts and in signal voids in the surrounding of instruments and especially near their tips. The artifacts generated by the susceptibility of the material are not only dependent on the material properties themselves and on the applied MRI sequences and parameters, but also on the geometric shape of the instruments and on the orientation to the static magnetic field in the MR unit. A numerical model based on superposition of induced elementary dipole fields was developed for studying the field distortions near paramagnetic needle tips. The model was validated by comparison with experimental data using field mapping MRI techniques. Comparison between experimental data and numerical simulations revealed good correspondence for the induced field inhomogeneities. Further systematic numerical studies of the field distribution were performed for variable types of concentric and asymmetric tip shapes, for different ratios between tip length and needle diameter, and for different orientations of the needle axis in the external static magnetic field. Based on the computed local inhomogeneities of the magnetic field in the surroundings of the needle tips, signal voids in usual gradient echo images were simulated for a prediction of the artifacts. The practically relevant spatial relation between those artifacts and the hidden tip of the needle was calculated for the different tip shapes and orientations in the external field. As needle tip determination is crucial in interventional procedures, e.g., in taking biopsies, the present model can help to instruct the physician prior to surgical interventions in better estimating the needle tip position for different orientations and needle tip shapes as they appear in interventional procedures. As manufacturing prototypes with subsequent measurements of artifacts in MRI are a costly procedure the presented model may also help to optimize shapes of needle tips and of other parts of MR-compatible instruments and implants with low expense prior to production if some shape parameters can be chosen freely.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Bernd Müller-Bierl, Hansjörg Graf, Ulrike Lauer, Günter Steidle, Fritz Schick. 2004. Numerical modeling of needle tip artifacts in MR gradient echo imaging.. https://doi.org/10.1118/1.1640971

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2 µm × 2 µm resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗